FDA Recall Open, Classified

Incisive CT-Computed Tomography X-Ray System Model: 728143

Recall: Z-2258-2023 · Initiated April 25, 2023

Recall

Recall Number
Z-2258-2023
Event Number
92202
Firm
Philips North America
FEI Number
3006648320
Product Code
JAK
Status
Open, Classified
Root Cause
Under Investigation by firm
Initiated
April 25, 2023
Posted
July 28, 2023
Address
222 Jacobs St, Cambridge, MA, 02141-2296

Description

Incisive CT-Computed Tomography X-Ray System Model: 728143

Reason

Metal mounting box on the rotating scanner on rotor (heat exchanger box) located within the Incisive CT system may become expelled and make contact with other components and cause injury to user or if detached a loud noise will be emitted and the system will shut down

Action

Phiips issued Urgent Medical Device Correction Letters mailed to US consignees via Certified United States Postal Service (USPS) with delivery confirmation on 4/25/23 International global markets were notified of the issue and instructed to distribute the FSN to affected consignees in accordance with local requirements. Letter states reason for recall, health risk and action to take: Please continue to use your system in accordance with its intended use. If you need to remain in the room for the duration of a scan, Philips recommends using protective eyewear. " Circulate this Urgent Medical Device Correction Letter to all users of this device so that they are aware of the issue. Please retain this letter with your system(s) until a solution is installed on your system; ensure the letter is in a place likely to be seen/viewed. " Place this URGENT Medical Device Correction Letter with your system documentation. " Please complete and return the attached response form to Philips promptly and no later than 30 days from receipt. Completing this form confirms receipt of the Urgent Medical Device Correction Letter, understanding of the issue, and required actions to be taken. Philips will contact you to schedule a time for a Field Service Engineer (FSE) to visit your site and install a solution to address the issue (reference FCO72800798). If you need additional information or support concerning this issue, please contact the Customer Care Solutions Center (1-800-722-9377).

Distribution

Nationwide Foreign: China Algeria Argentina Australia Austria Azerbaijan Bahrain Bangladesh Belarus Belgium Bolivia Brazil Bulgaria Canada Chile China Colombia Czech Republic Ecuador Egypt France Germany Greece India Indonesia Iran Iraq Israel Italy Japan Jordan Korea, Republic of Kosovo Latvia Macedonia Mexico Moldova Netherlands Nigeria Pakistan Panama Philippines Poland Romania Russian Federation Saint Lucia Saint Pierre and Miquelon Saudi Arabia Serbia Slovakia Spain Sweden Taiwan Thailand Tunisia Turkey Turkmenistan United Kingdom Uzbekistan Viet Nam

Quantity

36 US; 610 OUS