FDA Recall Open, Classified

System, Tomography Computed Emmission

Recall: Z-2258-2020 · Initiated November 30, 2018

Recall

Recall Number
Z-2258-2020
Event Number
85746
Firm
GE Healthcare, LLC
FEI Number
2126677
Product Code
KPS
Status
Open, Classified
Root Cause
Other
Initiated
November 30, 2018
Posted
June 3, 2020
Address
3000 N Grandview Blvd, Waukesha, WI, 53188-1615

Description

System, Tomography Computed Emmission

Reason

GE Healthcare has identified that a small number of Nuclear Medicine systems are operating without currently updated software. In a recent event, a patient s finger was injured during the Unload process while the automatic Unload feature was being used; the software on the system was not current.

Action

GE Healthcare identified that a small number of Nuclear Medicine systems are operating without currently updated software.

Distribution

Domestic Distribution: AL, AR, AZ, CA, CO, CT, DC, DE, FL,GA, IA, IL, IN, KY, LA,MA, MD, ME, MI, MN, MO, MS, NC, NE, NM, NJ, NV, NY, OH, OK, OR, PA, RI, SC, TN, TX, UT,VA, WI and WV. Foreign Distribution: Argentina, Australia, Belgium, Brazil, Canada, Colombia, Denmark, France, Germany, Ghana, India, Israel, Italy, Japan, Korea, Malaisya, Mexico, Netherlands, New Zealand, Norway, Poland, Portugal, Singapore, Sweden, Taiwan, Thailand, Turkey and United Kingdom.

Quantity

526 total units