FDA Recall
Open, Classified
Fusion Lithotripsy Extraction Basket, Model Number FS-LXB-2X4 G48277; Fusion Lithotripsy Extraction Baskets are intended to be used for endoscopic removal of biliary stones and foreign bodies.
Recall: Z-2252-2023
·
Initiated June 22, 2023
Recall
- Recall Number
- Z-2252-2023
- Event Number
- 92648
- Firm
- Wilson-Cook Medical Inc.
- FEI Number
- 1037905
- Product Code
- FFL
- Status
- Open, Classified
- Root Cause
- Process control
- Initiated
- June 22, 2023
- Posted
- July 28, 2023
- Address
- 4900 Bethania Station Rd, Winston Salem, NC, 27105-1203
Description
Fusion Lithotripsy Extraction Basket, Model Number FS-LXB-2X4 G48277; Fusion Lithotripsy Extraction Baskets are intended to be used for endoscopic removal of biliary stones and foreign bodies.
Reason
Non-conforming product may appear in the field as inability to advance or retract the basket, or the drive wire breaking or separating from handle (with or without mechanical lithotripsy).
Action
Cook Medical issued an URGENT: MEDICAL DEVICE REMOVAL notice to their consignees on 07/14/2023 by email or priority mail. The notice explained the issue, the risk, and requested the consignee cease use and return the affected product.
Distribution
Worldwide distribution.
Quantity
393 units