FDA Recall
Terminated
Philips IntelliSpace PACS 4.4 image management systems performs digital image processing, measurement and communication and storage.
Recall: Z-2252-2014
·
Initiated July 24, 2014
Recall
- Recall Number
- Z-2252-2014
- Event Number
- 68979
- Firm
- Philips Healthcare Informatics, Inc.
- FEI Number
- 3002971142
- Product Code
- LLZ
- Status
- Terminated
- Root Cause
- Software design
- Initiated
- July 24, 2014
- Posted
- August 16, 2014
- Terminated
- February 11, 2016
- Address
- 4100 E 3rd Ave, Ste 101, Foster City, CA, 94404-4819
Description
Philips IntelliSpace PACS 4.4 image management systems performs digital image processing, measurement and communication and storage.
Reason
A problem was detected in the Philips IntelliSpace PAC 4.4 software where if patient information is changed in the DICOM header and the study is exported before the image is updated in the PACS, the unchanged data may export causing the potential for diagnosis on incomplete data.
Action
Urgent Medical Device Correction was sent to all affected sites on July 24, 2014 by certified mail.
Distribution
US and Australia, Belgium, Canada, Egypt, France, Germany, Netherlands, Oman, Saudi Arabia, Spain, Switzerland, United Arab Emirates, United Kingdom.
Quantity
109 units