Fabius GS Tiro Anesthesia Machine; Cat. no. 8606000 Product Usage: Inhalation anesthesia machines for use in operating, induction and recovery rooms. They may be used with O2, N2O, and AIR supplied by a medical gas pipeline system or by externally mounted gas cylinders. Fabius series machines are equipped with a compact breathing system; providing fresh gas decoupling, PEEP, and pressure limitation. The following ventilation options are available: Volume Controlled Ventilation Pressure Controlled Ventilation Pressure Support (Optional) SIMV/PS (Optional) Manual Ventilation Spontaneous Breathing Fabius series anesthesia machines are equipped with an electrically driven and electronically controlled ventilator and monitors for airway pressure (P), volume (V), and inspiratory oxygen concentration (FiO2). As per IEC 60601-2-13 (Anesthetic Workstations and their Modules-Particular Requirements), additional monitoring of the concentrations of CO2 and anesthetic agent is required when the machine is in use. IEC 60601-2-13:2003 requires that a manual ventilation bag must be available for emergency use. Fresh gas enrichment is provided by the Drger Vapor anesthetic vaporizer. Fabius MRI can also be used in MRI scanner rooms with magnets 1.5 tesla and 3 tesla at a fringe field strength of 40 mtesla or less
Recall
- Recall Number
- Z-2251-2018
- Event Number
- 80209
- Firm
- Draeger Medical, Inc.
- FEI Number
- 2517967
- Product Code
- BSZ
- Status
- Terminated
- Root Cause
- Process change control
- Initiated
- May 24, 2018
- Terminated
- May 12, 2020
- Address
- 3135 Quarry Rd, Telford, PA, 18969-1042
Description
Fabius GS Tiro Anesthesia Machine; Cat. no. 8606000 Product Usage: Inhalation anesthesia machines for use in operating, induction and recovery rooms. They may be used with O2, N2O, and AIR supplied by a medical gas pipeline system or by externally mounted gas cylinders. Fabius series machines are equipped with a compact breathing system; providing fresh gas decoupling, PEEP, and pressure limitation. The following ventilation options are available: Volume Controlled Ventilation Pressure Controlled Ventilation Pressure Support (Optional) SIMV/PS (Optional) Manual Ventilation Spontaneous Breathing Fabius series anesthesia machines are equipped with an electrically driven and electronically controlled ventilator and monitors for airway pressure (P), volume (V), and inspiratory oxygen concentration (FiO2). As per IEC 60601-2-13 (Anesthetic Workstations and their Modules-Particular Requirements), additional monitoring of the concentrations of CO2 and anesthetic agent is required when the machine is in use. IEC 60601-2-13:2003 requires that a manual ventilation bag must be available for emergency use. Fresh gas enrichment is provided by the Drger Vapor anesthetic vaporizer. Fabius MRI can also be used in MRI scanner rooms with magnets 1.5 tesla and 3 tesla at a fringe field strength of 40 mtesla or less
The Draeger anesthesia device may be able to dose 100% N2O. In the event of a fault, the S-ORC module would not prevent setting an N2O flow that would result in a hypoxic mixture from being dosed to the patient. Potential adverse outcomes include death of the patient.
On May 24, 2018, the firm issued Urgent Medical Device Recall letters to affected end users (medical facilities) via certified mail. The letter identified the affected product, problem and actions to be taken. The recalling firm stated that they are currently manufacturing replacement parts for all potentially affected devices, which will be installed as part of the next service. The firm's Service dept. will contact end users to arrange the replacement Free of Charge In addition, end users were asked to complete and return a Customer Reply Card to acknowledge the receipt of this information. If you have any questions regarding this recall please contact Michael Kelhart between the hours of 8:00 AM 4:30 PM EST at 1-800-437-2437 (press 1 at the prompt, then 32349). For questions regarding the operation and/or servicing of your Draeger anesthesia machine please contact DraegerService Technical Support between the hours of 8:00 AM 8:00 PM EST at 1-800-437-2437 (press 4 at the prompt then 4 again).
US Nationwide Distribution - Healthcare facilities in AK, AL, AZ, CA, CO, FL, GA, IA, IL, IN, MA, MD, MN, MO, MS, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, TN, TX, UT, WA, WI, and WV.
31