FDA Recall Open, Classified

Carina Sub-Acute Care Ventilator

Recall: Z-2248-2023 · Initiated July 12, 2023

Recall

Recall Number
Z-2248-2023
Event Number
92629
Firm
Draeger Medical, Inc.
FEI Number
2517967
Product Code
CBK
Status
Open, Classified
Root Cause
Nonconforming Material/Component
Initiated
July 12, 2023
Posted
August 18, 2023
Address
3135 Quarry Rd, Telford, PA, 18969-1042

Description

Carina Sub-Acute Care Ventilator

Reason

Presence of polyether polyurethane (PE-PUR) and emission of 1,3-Dichloropropan-2-ol that exceed the acceptable uptake level during continuous use (>30 days) in pediatric patients.

Action

The firm, Drager, sent an "Urgent Medical Device Recall" to customers by mail on 07/12/2023. The medical device recall notification instructs customers on how to utilize the Carina Ventilator until units can be corrected in the field by way of replacement of the blower to remove the presence of PE-PUR foam. Customers may continue to use the device until it has been modified provided that the set minute volume exceeds 3.6 l/min, only adult patients are ventilated, and only a Draeger ventilation hose with leak valve is used. The firm states that a new designed blower cover will be available in the early part of the fourth quarter of 2023 and Draeger will facilitate field service free of charge. Customers are to ensure all users and maintenance staff that utilize the recalled device are made aware of the recall notice. Customers with any questions regarding the operation of the Carina Ventilator are to contact Technical Support between 8:00 AM and 8:00 PM EST at 1-800-437-2437 (press 2, then 2, then 2); customers with any questions about the recall notice are to contact Michael Kelhart from 8:00 AM to 4:30 PM EST at 267-664-1131 or by email at [email protected]. The Medical Device Recall Response form is to be returned by fax 215-372-2940 or email: [email protected] to Draeger upon completion.

Distribution

US Nationwide Distribution.

Quantity

11,621 units