FDA Recall Open, Classified

GE HealthCare ApexPro v6.0 eDelivery Software, REF #5700026, Product Code SUV; used in conjunction wtih ApexPro CARESCAPE Telemetry Server (CTS) v6.0; Monitor, Physiological, Patient(With Arrhythmia Detection Or Alarms)

Recall: Z-2247-2026 · Initiated April 24, 2026

Recall

Recall Number
Z-2247-2026
Event Number
98938
Firm
GE Medical Systems Information Technologies Inc
FEI Number
3001570235
Product Code
MHX
Status
Open, Classified
Root Cause
Software design
Initiated
April 24, 2026
Posted
May 28, 2026
Address
3114 N Grandview Blvd, Waukesha, WI, 53188-1615

Description

GE HealthCare ApexPro v6.0 eDelivery Software, REF #5700026, Product Code SUV; used in conjunction wtih ApexPro CARESCAPE Telemetry Server (CTS) v6.0; Monitor, Physiological, Patient(With Arrhythmia Detection Or Alarms)

Reason

GE HealthCare has become aware of an unlikely situation where potential loss of ECG and SpO2 monitoring involving ApexPro CARESCAPE Telemetry Server (CTS) v5.0 and v6.0, can result from a persistent NO COMM or OFF NETWORK condition at the CARESCAPE Central Station or CIC Pro Clinical Information Center (Central Station).

Action

GE HealthCare issued an URGENT MEDICAL DEVICE CORRECTION notice to its consignees on 4/24/2026 by letter using a traceable means. The notice explained the the safety issue pertaining to the persistent NO COMM or OFF NETWORK condition, and provided instructions for the safe use of the device should either issue arise. "Actions to be taken by Customer/User Pending corrections from GE HealthCare, you can continue to use your ApexPro CTS with Central Station for telemetry only or combination monitoring unless you observe a persistent (more than a few seconds) NO COMM or OFF NETWORK notification. If you do observe a persistent NO COMM or OFF NETWORK notification, follow the instructions below. 1. Use alternative bedside patient monitoring to ensure continuous clinical surveillance until the NO COMM or OFF NETWORK condition is resolved by GE HealthCare. NOTE: If you are using combination monitoring, it is recommended to switch ECG monitoring to the bedside monitor. 2. Immediately notify your Facility s Biomedical Engineering, IT or Service Department and contact GE HealthCare Service. 3. Verify monitoring is restored at the Central Station(s) and/or bedside monitor(s). Please ensure all potential users in your facility are made aware of this safety notification and the recommended actions. Please post this notification by your Central Station(s). Please retain this document for your records." "GE HealthCare will correct all affected products at no cost to you. A GE HealthCare representative will contact you to arrange for the correction." For questions or concerns regarding this notification, please contact GE HealthCare Service at 1-800-437-1171 or your local Service Representative.

Distribution

Worldwide distribution.

Quantity

2187 units