FDA Recall Terminated

Part no. 690.380, Graphic Case F/7.3mm Cannulated Screw Set

Recall: Z-2247-2018 · Initiated March 5, 2018

Recall

Recall Number
Z-2247-2018
Event Number
79754
Firm
Synthes (USA) Products LLC
FEI Number
3008812560
Product Code
FSM
Status
Terminated
Root Cause
Process control
Initiated
March 5, 2018
Terminated
February 28, 2020
Address
1301 Goshen Pkwy, West Chester, PA, 19380-5986

Description

Part no. 690.380, Graphic Case F/7.3mm Cannulated Screw Set

Reason

Affected screw gauges incorrectly measure screws 5.41mm shorter than the actual overall screw length.

Action

On March 8, 2018, the recalling firm sent Urgent Product Recall letters to its customers to inform them of the recall. Customers were advised to take the following actions: 1. Immediately review their inventory to identify and quarantine all affected products listed above in a manner that ensures the affected products will not be used. 2. Complete the Verification Section and send a copy to the recalling firm. 3. Forward this notice to anyone in the facility that needs to be informed. 4. If any of the affected products have been forwarded to another facility, contact that facility to arrange return. Customers with questions were advised to call 610-719-5450, 8AM-5PM (EST), Monday through Friday, or contact their DePuy Synthes sales consultant.

Distribution

Distributed to accounts in CA, IL, IN, KS, LA, MD, NC, NE, NJ, OR, TX, and VA. Foreign distribution to Canada.

Quantity

3