FDA Recall Open, Classified

First Source Vision M. Portable cart and platform that supports an X-ray system and laptop computer.

Recall: Z-2246-2024 · Initiated May 9, 2024

Recall

Recall Number
Z-2246-2024
Event Number
94744
Firm
First Source Inc
FEI Number
3004063527
Product Code
IZL
Status
Open, Classified
Root Cause
Device Design
Initiated
May 9, 2024
Posted
July 1, 2024
Address
3495 Winton Pl, Bldg E Ste 1, Rochester, NY, 14623-2824

Description

First Source Vision M. Portable cart and platform that supports an X-ray system and laptop computer.

Reason

Potential for the bolt that secures the positioning arm to the gas spring to come loose and fall onto patient.

Action

Consignees were emailed an URGENT: MEDICAL DEVICE RECALL notification, dated 5/29/24. The notice instructs consignees to follow the recommended preventive maintenance schedule and visually check the device before use to identify if the affected bolt is coming loose. The recalling firm is currently testing a spring clip which will keep the bolt from falling out and will send this part to customers at no charge once approved. Consignees are asked to return the provided response form. Consignees with any questions can contact the firm by email at [email protected] or by phone at 1-800-349-5980 (option 2) from 9:00 AM to 5:00 PM EST from Monday through Friday.

Distribution

Worldwide - US Nationwide distribution in the states of AR, CA, FL, GA, IL, IN, LA, MA, MD, MI, MO, MS, NC, NY, OK, PA, TN, TX, & VA. The countries of Korea & Serbia.

Quantity

680 units