FDA Recall Terminated

FilmArray Blood Culture Identification (BCID) Panel Product Pouch Label: FilmArray BCID Panel Part No: RFIT-ASY-0126 Part No: RFIT-ASY-0127

Recall: Z-2245-2021 · Initiated July 20, 2021

Recall

Recall Number
Z-2245-2021
Event Number
88410
Firm
BioFire Diagnostics, LLC
FEI Number
3002773840
Product Code
PEN
Status
Terminated
Root Cause
Process control
Initiated
July 20, 2021
Terminated
December 16, 2022
Address
515 S Colorow Dr, Salt Lake City, UT, 84108-1248

Description

FilmArray Blood Culture Identification (BCID) Panel Product Pouch Label: FilmArray BCID Panel Part No: RFIT-ASY-0126 Part No: RFIT-ASY-0127

Reason

Due to product complaints received for false negative results (dropouts) with blood culture identification panel potential caused by a pouch manufacturing error.

Action

On 07/20/2021, the firm sent an "URGENT: MEDICAL DEVICE CORRECTION" Notification Letter to customer informing them that an issue has been identified in the manufacturing process which might contribute to elevated rates of false negative results while using the BCID Panel. Actions to be taken by customers: 1) Immediately examine their inventory for product identified in this recall, 2) Discontinue use and discard any remaining product in their possession that is subject to this recall. The Recalling Firm will replace the product at no charge in accordance with its standard limited warranty. 3) Complete the accompanying Field Action Acknowledgment of Receipt Form and confirm the number of pouches scrapped (if any). Return the Acknowledgment of Receipt to the Recalling Firm so that it may acknowledge their receipt of this notification. 4) If they may have further distributed the affected product, identify any recipients and notify them at once of this product recall. For product replacement, as well as any other questions or concerns, contact customer support department at [email protected] or via telephone at 1-800-735-6544 and selecting option 5 for Product Technical Support.

Distribution

U.S. Nationwide Distribution in the states of : FL, MI, NJ, OH, VA and WI. O.U.S.: None

Quantity

29 kits