FDA Recall
Terminated
Medtronic O-arm Imaging System. Product Numbers: BI-700-00027-100, BI-700-00027-100R, BI-700-00027-120, BI-700-00027-120R, BI-700-00027-230, BI-700-00027-230R.
Recall: Z-2244-2012
·
Initiated July 11, 2012
Recall
- Recall Number
- Z-2244-2012
- Event Number
- 62571
- Firm
- Medtronic Navigation, Inc.
- FEI Number
- 3004785967
- Product Code
- OXO
- Status
- Terminated
- Root Cause
- Component change control
- Initiated
- July 11, 2012
- Posted
- August 22, 2012
- Terminated
- June 6, 2016
- Address
- 300 Foster St, Littleton, MA, 01460-2017
Description
Medtronic O-arm Imaging System. Product Numbers: BI-700-00027-100, BI-700-00027-100R, BI-700-00027-120, BI-700-00027-120R, BI-700-00027-230, BI-700-00027-230R.
Reason
A fault in a high voltage relay in the 0-arm Imaging System might cause the relay to overheat and burn.
Action
Medtronic Navigation issued an "Urgent Field Safety Notice" to consignees dated 7/11/12. The problem was described and recommended actions were provided. For each affected O-arm Imaging System, a Medtronic service representative will visit the site and replace the 500VDC relay on the system and test the O-arm Imaging System.
Distribution
Worldwide Distribution - USA, including the states of AL, CO, FL, IA, MI, IA, NC, NE, NY and the countries of South Africa, Italy, and Spain.
Quantity
17 units