FDA Recall Terminated

EasyLink Informatic System is included with the Dimension Vista System. The Dimension Vista Integrated System is an in vitro diagnostic device intended tot duplicate manual analytical procedures such as pipetting, mixing, heating, and measuring spectral intensities to determine a variety of analytes in human body fluids. The EasyLink Informatics System is classified as a data calculator and processing module for use with in vitro diagnostic devices. The EasyLink is included with the Dimension Vista System and may be used with other Siemens analyzers as a communications and connectivity workstation for integration with laboratory information system (LIS) networks.

Recall: Z-2242-2012 · Initiated May 29, 2012

Recall

Recall Number
Z-2242-2012
Event Number
62460
Firm
Siemens Healthcare Diagnostics, Inc.
FEI Number
2517506
Product Code
JQP
Status
Terminated
Root Cause
Software design
Initiated
May 29, 2012
Posted
August 22, 2012
Terminated
April 10, 2014
Address
500 GBC Drive PO BOX 6101, Mailstop 514, Newark, DE, 19714-6101

Description

EasyLink Informatic System is included with the Dimension Vista System. The Dimension Vista Integrated System is an in vitro diagnostic device intended tot duplicate manual analytical procedures such as pipetting, mixing, heating, and measuring spectral intensities to determine a variety of analytes in human body fluids. The EasyLink Informatics System is classified as a data calculator and processing module for use with in vitro diagnostic devices. The EasyLink is included with the Dimension Vista System and may be used with other Siemens analyzers as a communications and connectivity workstation for integration with laboratory information system (LIS) networks.

Reason

When using the EasyLink Quality Control Package the system may not perform as intended for Multi-Rule QC violations [e.g., 2(2S), 4(1s) and 10x], unless configured using the instructions send with the Urgent Field Safety Notice. When this situation occurs, QC status may not be flagged appropriately. If auto-verification rules are defined to hold patient results when QC status is out, the patient

Action

Siemens sent an "URGENT FIELD SAFETY NOTICE" dated May 2012 to all affected customers. The letter identifies the product, problem, and actions to be taken by the customers. A Field Correction Effectiveness Check Form was attached for customers to complete and return via fax to 302-631-8467. Contact Siemens Technical Solutions Center at 800-441-9250 for questions regarding this notice.

Distribution

Worldwide Distribution-USA (nationwide) and the countries of Canada, France, Germany, Spain, Portugal, Italy, Japan, Norway, Australia, Austria, New Zealand, Belgium, Switzerland, Netherlands, Malaysia, Slovenia, South Korea, Denmark, Saudi Arabia, Slovakia, Taiwan, Philippines, Czech Republic, and Singapore.

Quantity

637