CIVCO Surgi-Tip Transducer Cover, Ref. 610-834, Sterile (20.3 x 244cm) polyethylene telescopically-folded cover with attached (6cm) NeoGuard tip. The latex-free surgi tipped covers are utilized to cover ultrasound transducers and prevent cross-contamination from occurring with equipment and patients.
Recall
- Recall Number
- Z-2242-2010
- Event Number
- 56285
- Firm
- Civco Medical Instruments Inc
- FEI Number
- 1937223
- Product Code
- ITX
- Status
- Terminated
- Root Cause
- Nonconforming Material/Component
- Initiated
- July 20, 2010
- Posted
- August 20, 2010
- Terminated
- April 13, 2012
- Address
- 102 1st St S, Kalona, IA, 52247-9589
Description
CIVCO Surgi-Tip Transducer Cover, Ref. 610-834, Sterile (20.3 x 244cm) polyethylene telescopically-folded cover with attached (6cm) NeoGuard tip. The latex-free surgi tipped covers are utilized to cover ultrasound transducers and prevent cross-contamination from occurring with equipment and patients.
Polyethylene material did not have proper treatment applied during manufacturing. Untreated polyethylene material does not maintain a secure seal with the adhesive tape used to secure the tip to the cover.
CIVCO sent a "Urgent - Medical Device Correctdion" letter date July 19, 2010 to consignees describing the reason for recall and requesting they dispose of the affect lot number of product. A response form was enclosed for completion and return to the firm, indicating the amount of product in inventory and requesting a Return Authorization Number. The form was to be returned via FAX or e-mail. Customers may contact CIVO at 800-445-6741.
Nationwide distribution: the states of OH, MD, NE, VA, IA, MI, WA, and WI.
22 boxes