FDA Recall Terminated

Dimension Vista Homocysteine Flex Reagent Cartridge (HCYS), Device Listing No.: D017878 is an in vitro diagnostic test for the quantitative measurement of total homocysteine in human serum, heparinized plasma and EDTA plasma on the Dimension Vista System.

Recall: Z-2241-2016 · Initiated May 17, 2016

Recall

Recall Number
Z-2241-2016
Event Number
74468
Firm
CSL Behring GmbH Emil-von-Behring-Str. 76 Marburg Germany
FEI Number
3003098680
Product Code
LPS
Status
Terminated
Root Cause
Under Investigation by firm
Initiated
May 17, 2016
Terminated
February 7, 2018

Description

Dimension Vista Homocysteine Flex Reagent Cartridge (HCYS), Device Listing No.: D017878 is an in vitro diagnostic test for the quantitative measurement of total homocysteine in human serum, heparinized plasma and EDTA plasma on the Dimension Vista System.

Reason

Reagent lots # 15243MA and 15243MB shows a reduced stability once opened that does not meet the Instructions For Use claim of seven (7) days. A maximum bias of -37% was observed.

Action

Siemens mailed an Urgent Medical Device Recall to all customers who received the affected lots notifying them of the following: 1) Discontinue use and discard any remaining inventory of the affected lots, and 2) Indicate their replacement product needs within 30 days.

Distribution

Distributed to: AL, AR, CA, CO, CT, DC, DE, FL, GA, IA, IL, IN, LA, MA, MD, MI, MO, MS, MT, NC, NE, NH, NJ, NV, NY, OH, OR, PA, RI, SC, TN, TX, UT, VA, WI, WV

Quantity

898 units