FDA Recall Terminated

TSP Hip Fracture Plating System device, fixation, proximal femoral, implant The Biomet Hip Fracture Plating Systems are indicated for use in the open reduction and internal fixation of a wide variety of fractures of the proximal femur: intracapsular fractures and intertrochanteric fractures. Fracture plates with six or more femoral shaft fixation screw holes are also indicated for subtrochanteric fractures as well as proximal femoral osteotomies

Recall: Z-2240-2016 · Initiated June 10, 2016

Recall

Recall Number
Z-2240-2016
Event Number
74576
Firm
Biomet, Inc.
FEI Number
1825034
Product Code
JDO
Status
Terminated
Root Cause
Device Design
Initiated
June 10, 2016
Posted
July 19, 2016
Terminated
July 20, 2017
Address
56 E Bell Dr, Warsaw, IN, 46582-6989

Description

TSP Hip Fracture Plating System device, fixation, proximal femoral, implant The Biomet Hip Fracture Plating Systems are indicated for use in the open reduction and internal fixation of a wide variety of fractures of the proximal femur: intracapsular fractures and intertrochanteric fractures. Fracture plates with six or more femoral shaft fixation screw holes are also indicated for subtrochanteric fractures as well as proximal femoral osteotomies

Reason

The screw may fracture at the junction of the shaft and the head during implantation. The patient may retain a foreign body if the fractured screw cannot be removed. Soft tissue irritation will not occur as the retained portion of the screw is below the surface of the plate. A delay in surgery less than 30 minutes may occur.

Action

Zimmer Biomet sent an Urgent Medical Device Recall letter dated June 10, 2016, to all affected customers via courier. Distributors and hospitals were instructed to locate and return the affected product, and implanting surgeons were instructed to monitor patients. The recall notification included a description of the reason for the recall, affected product, consignee responsibilities, and instructions for responding to the formal recall notification. Questions related to this notice should be directed to (574) 372-6725, Monday through Friday, 8 a.m. to 5 p.m.

Distribution

Worldwide Distribution - US (Nationwide) Internationally to Australia, The Netherlands, and Japan

Quantity

5458 units