FDA Recall Open, Classified

Pinnacle 3 with TumorLOC, (870258) Radiation Therapy Planning System Software

Recall: Z-2239-2025 · Initiated July 17, 2025

Recall

Recall Number
Z-2239-2025
Event Number
97309
Firm
Philips Medical Systems (Cleveland) Inc
FEI Number
3018004899
Product Code
MUJ
Status
Open, Classified
Root Cause
Software change control
Initiated
July 17, 2025
Posted
August 1, 2025
Address
3630 Sw 47th Ave, Gainesville, FL, 32608-7555

Description

Pinnacle 3 with TumorLOC, (870258) Radiation Therapy Planning System Software

Reason

Due to software issue, Radiation Therapy Planning system may provide incorrect dataset calculations when performing the "Stopping Power Ratio" (SPR) ,

Action

On July 17, 2025, Philips issued a "Urgent Medical Device Correction" Notification to affected consignees. Philips asked consignees to take the following actions: 1. You may continue to use your system(s) in accordance with the intended use and by following the recommendations below. 2. To prevent the issue from occurring, when using SPR generation, always verify data sets are from same acquisition and matching reconstruction parameters. 3. Ensure that relevant clinical staff, including physicians, physicists, and dosimetrists, are informed about the potential for incorrect dataset calculations due to mixing of data sets 4D CT spectral and SPR generation. 4. Circulate this notice to all users of this device so that they are aware of the potential issue. Please retain this letter with your system(s) until a solution is installed on your system; ensure the letter is in a place likely to be seen/viewed. 5. Please complete and return the attached response form to Philips promptly and no later than 30 days from receipt of this letter via email to: [email protected]

Distribution

Worldwide - US Nationwide distribution in the state of WV and the countries of Macao & United Kingdom.

Quantity

149 Serial Numbers