Beckman Coulter ACCESS Immunoassay Systems, Access Total T3 Triiodothyronine , REF 33830 (UDI: 15099590227210) for use with the Access Family of Immunoassay Systems including: Access 2, UniCel DxI 600, UniCel DxI 800, UniCel DxC 600i, UniCel DxC 660i, UniCel DxC 680i, UniCel DxC 860i, and UniCel DxC 880i systems.

Recall: Z-2239-2020 · Initiated April 30, 2019

Recall

Recall Number
Z-2239-2020
Event Number
84862
Firm
Beckman Coulter Inc.
FEI Number
2050012
Product Code
CDP
Status
Terminated
Root Cause
Other
Initiated
April 30, 2019
Posted
June 2, 2020
Terminated
December 28, 2021
Address
250 S Kraemer Blvd, Brea, CA, 92821-6232

Description

Beckman Coulter ACCESS Immunoassay Systems, Access Total T3 Triiodothyronine , REF 33830 (UDI: 15099590227210) for use with the Access Family of Immunoassay Systems including: Access 2, UniCel DxI 600, UniCel DxI 800, UniCel DxC 600i, UniCel DxC 660i, UniCel DxC 680i, UniCel DxC 860i, and UniCel DxC 880i systems.

Reason

The recalling firm characterized the potential interference effects of biotin (up to 1,200 ng/mL) on all biotin-mediated Access assays. The following six assays: Free T3-REF A13422, Total T3-REF 33830, Free T4-REF 33880, GI Mon-387687, Tg-REF 33860, and TgAbII-REF A32898 displayed significant interference (defined as a result bias greater than +/-10%) at 1,200 ng/mL of biotin.

Action

On 04/30/2019, an Urgent Medical Device Letter was sent via mail to affected consignees. Customers were instructed to access the updated Instructions for Use (IFU) Limitations and Interferences sections for the six affected Access immunoassays on the firm's website. Additionally they ask customers to do the following: 1) Share the information with laboratory staff and retain the notification as part of their laboratory Quality System documentation. If any of the affected product(s) have been forwarded to another laboratory, they are to provide them with a copy of the Urgent Medical Device Recall letter. 2) Respond within 10 days either electronically (if received communication via email) or manually (complete and return an enclosed Response Form). If there are any questions regarding the notice, contact the Customer Support Center: website: http://www.beckmancoulter.com By phone: call 1-800-854-3633 in the United States and Canada. Outside the United States and Canada, contact the local firm representative.

Distribution

US Distribution: Alabama, Alaska, Arizona, Arkansas, California, Colorado, Connecticut, Delaware, District of Columbia, Florida, Georgia, Hawaii, Idaho, Illinois, Indiana, Iowa, Kansas, Kentucky, Louisiana, Maine, Maryland, Massachusett, Michigan, Minnesota, Mississippi, Missouri, Nebraska, Nevada, New Hampshire, New Jersey, New Mexico New York, North Carolina, North Dakota, Ohio, Oklahoma Oregon, Pennsylvania, Rhode Island, South Carolina,South Dakota, Tennessee, Texas, Utah, Vermont, Virginia, Washington, West Virginia, Wisconsin, and Wyoming. US Territories: American Samoa, Guam, and Puerto Rico. OUS Distribution: Algeria, Andorra, Australia, Austria, Bahrain, Bangladesh, Belgium, Bosnia and Herzegovina, Botswana, Brazil, Brunei, Bulgaria, Canada, Chile, China, Costa Rica, Croatia, Curacao, Cyprus, Czech Republic, Denmark, Djibouti, Ecuador, Egypt, El Salvador, Estonia, Ethiopia, Finland, France, Georgia, Germany, Ghana, Gibraltar, Greece, Guam, Guatemala, Guyana, Hong Kong, Hungary, India, Indonesia, Ireland, Israel, Italy, Japan, Jordan, Kazakhstan, Kenya, Kuwait, Lebanon, Liberia, Lithuania, Malawi, Malaysia, Mayotte, Mexico, Morocco, Mozambique, Myanmar, Namibia, Nepal, Netherlands, Netherlands Antillean, New Zealand, Nigeria, Oman, Pakistan, Panama, Paraguay, Peru, Philippines, Poland, Portugal, Qatar, Romania, Russia, Rwanda, San Marino, Serbia, Sierra Leone, Singapore, Slovakia, South Africa, South Korea, Spain, Sri Lanka, Swaziland, Sweden, Switzerland, Taiwan, Thailand, Tunisia, Turkey, Uganda, United Arab Emirates, United Kingdom, Uruguay, Venezuela, Vietnam, Zambia, and Zimbabwe.

Quantity

238,862