FDA Recall Terminated

GE Inspection Technologies, LP - Phoenix v|tome|x m240 cabinet x-ray systems. A microfocus CT system used primarily for 3D metrology and analysis.

Recall: Z-2236-2016 · Initiated March 2, 2016

Recall

Recall Number
Z-2236-2016
Event Number
74554
FEI Number
2519904
Product Code
RCE
Status
Terminated
Root Cause
Radiation Control for Health and Safety Act
Initiated
March 2, 2016
Posted
July 23, 2016
Terminated
August 7, 2017
Address
Rd, GE Inspection Technologies, LP 50 Industrial Park, Lewistown, PA, 17044-9312

Description

GE Inspection Technologies, LP - Phoenix v|tome|x m240 cabinet x-ray systems. A microfocus CT system used primarily for 3D metrology and analysis.

Reason

Radiation emissions which exceeded the limit in the performance standard were detected during maintenance of one system. The identified cause was insufficient attachment of shielding in the door. Jostling during delivery was identified as the probable cause of this failure to comply with the performance standard.

Action

GE plans to notify its customers with the systems through a TIL. The TIL will inform customers that GE will add lead strips to the inside window frame under the lead glass window of the main cabinet door. GE recommends that it visit each customer site and repair the system by adding lead strips along the window frame of the cabinet door, beneath the lead glass window, within 90 days following FDA's approval of the repair plan.

Distribution

US Distribution to the states of : FL, TN, and TX.

Quantity

4