FDA Recall Open, Classified

vRad PACS with Mammography

Recall: Z-2235-2023 · Initiated May 9, 2023

Recall

Recall Number
Z-2235-2023
Event Number
92516
Firm
Virtual Radiologic Corp.
FEI Number
3007795813
Product Code
LLZ
Status
Open, Classified
Root Cause
Software Manufacturing/Software Deployment
Initiated
May 9, 2023
Posted
July 21, 2023
Address
11995 Singletree Ln, Ste 500, Eden Prairie, MN, 55344-5349

Description

vRad PACS with Mammography

Reason

The error resulted in intermittent failure of current (primary) radiology study images not displaying to the interpreting teleradiologist.

Action

The firm issued an URGENT FIELD CORRECION notice to its consignees beginning on 06/28/2023 via email. The notice explained the issue, root cause, and repair of the software. It was recommended that those consignees who experienced the malfunction review the affected cases.

Distribution

US Nationwide distribution in the states of MD, MN, TN, FL, GA, TX, OK.

Quantity

8 units