FDA Recall
Open, Classified
vRad PACS with Mammography
Recall: Z-2235-2023
·
Initiated May 9, 2023
Recall
- Recall Number
- Z-2235-2023
- Event Number
- 92516
- Firm
- Virtual Radiologic Corp.
- FEI Number
- 3007795813
- Product Code
- LLZ
- Status
- Open, Classified
- Root Cause
- Software Manufacturing/Software Deployment
- Initiated
- May 9, 2023
- Posted
- July 21, 2023
- Address
- 11995 Singletree Ln, Ste 500, Eden Prairie, MN, 55344-5349
Description
vRad PACS with Mammography
Reason
The error resulted in intermittent failure of current (primary) radiology study images not displaying to the interpreting teleradiologist.
Action
The firm issued an URGENT FIELD CORRECION notice to its consignees beginning on 06/28/2023 via email. The notice explained the issue, root cause, and repair of the software. It was recommended that those consignees who experienced the malfunction review the affected cases.
Distribution
US Nationwide distribution in the states of MD, MN, TN, FL, GA, TX, OK.
Quantity
8 units