FDA Recall Terminated

GUIDE WIRE w/ BALL TIP 2.0MMX98CM; 469040 Intended to be used as a guide pin for insertion of implants (cannulated screws and/or intramedullary nails). Bone nail guidewire. Intended for drilling bone.

Recall: Z-2234-2018 · Initiated May 16, 2018

Recall

Recall Number
Z-2234-2018
Event Number
80193
Firm
Zimmer Biomet, Inc.
FEI Number
1825034
Product Code
HTY
Status
Terminated
Root Cause
Under Investigation by firm
Initiated
May 16, 2018
Posted
June 1, 2018
Terminated
April 5, 2024
Address
56 E Bell Dr, Warsaw, IN, 46582-6989

Description

GUIDE WIRE w/ BALL TIP 2.0MMX98CM; 469040 Intended to be used as a guide pin for insertion of implants (cannulated screws and/or intramedullary nails). Bone nail guidewire. Intended for drilling bone.

Reason

The firm is recalling various trauma guide wires due to insufficient packaging design verification and exceeding the expected occurrence rate of complaints for sterile barrier failure. Guide wires packaged in the affected packaging configuration have the potential for sterile barrier breach, leading to possible risk for infection.

Action

The firm, Zimmer Biomet, sent an "URGENT MEDICAL DEVICE RECALL" letter dated May 16, 2018 on May 16, 2018, to affected consignees via email and FedEx. The letter describes the product, problem and actions to be taken. Consignees were informed of the recall and instructed to take the following actions: Risk Manager Responsibilities: 1. Review this notification and ensure that affected personnel are aware of the contents. 2. If you have affected product at your facility, assist your Zimmer Biomet sales representative and quarantine all affected product. Your Zimmer Biomet sales representative will remove the affected product from your facility. 3. Complete Attachment 1 Certificate of Acknowledgement and send to [email protected]. This form must be returned even if you do not have affected products at your facility. Distributor Responsibilities: 1. Review this notification and ensure that affected team members are aware of the contents. 2. Immediately locate and quarantine affected product in your inventory. 3. Immediately return all affected product from your distributorship and from affected hospitals within your territory. a. For each return, send a copy of Attachment 1 to [email protected] b. International returns, please request an IRA by emailing [email protected] c. Include a hardcopy of Attachment 1 in each carton of your return shipment for immediate processing. d. Mark RECALL on the outside of the returned cartons. 4. Return the Additional Accounts form to [email protected]. a. Review the list of hospitals included with the email notification sent to your facility, which includes a list of hospitals that have already been notified of this recall. b. Identify whether there are any additional hospitals that Zimmer Biomet has not notified and list these accounts on the Additional Accounts form. Please provide the form in Excel format

Distribution

Worldwide Distribution: US (Nationwide) and countries of: Argentina, South Korea, Chile, Colombia, Trinidad and Tobago, El Salvador, Costa Rica, Canada, Australia, Japan, and Netherlands.

Quantity

449