FDA Recall Terminated

Siemens brand Ultrasound Transducer 17L5HD used with ACUSON Sequoia Ultrasound System; Model Number(s): 10035724, used with System Models: 08245875, 08246951, 08255412, 08255413, 08267697, 08269627, 10038241, 10038242, 10040724, 10040725, 10041008; Product is manufactured and distributed by Siemens Medical Solutions USA, Inc. Ultrasound Division, Mountain View, CA The 17L5 transducer is intended for breast and small body parts imaging.

Recall: Z-2230-2008 · Initiated June 6, 2008

Recall

Recall Number
Z-2230-2008
Event Number
48695
Firm
Siemens Medical Solutions USA, Inc.
FEI Number
2936884
Product Code
IYO
Status
Terminated
Root Cause
Process control
Initiated
June 6, 2008
Posted
September 11, 2008
Terminated
December 21, 2010
Address
1230 Shorebire Way P.O. Box 7393, Mountain View, CA, 94043

Description

Siemens brand Ultrasound Transducer 17L5HD used with ACUSON Sequoia Ultrasound System; Model Number(s): 10035724, used with System Models: 08245875, 08246951, 08255412, 08255413, 08267697, 08269627, 10038241, 10038242, 10040724, 10040725, 10041008; Product is manufactured and distributed by Siemens Medical Solutions USA, Inc. Ultrasound Division, Mountain View, CA The 17L5 transducer is intended for breast and small body parts imaging.

Reason

The product assembly can cause a failure of the transducer which results in a double and/or overlapped image which may ultimately result in misleading or false information, inability to accurately diagnose, incorrect positioning/locating/insertion of the biopsy needle and/or an interruption during the biopsy procedure.

Action

On 6/6/08, the firm initiated the recall and its notification was via letters explaining the reason for the recall and requesting the affected product be returned. Affected transducers will be replaced. For assistance, contact Siemens Medical Solutions at 1-650-694-5993.

Distribution

Worldwide Distribution: USA, AUSTRALIA, AUSTRIA, CANADA, CHINA, DENMARK, FRANCE, GERMANY, INDIA, ITALY, JAPAN, KOREA, MALAYSIA, MEXICO, NORWAY, PORTUGAL, RUSSIAN FEDERATION, SINGAPORE, SWEDEN, TAIWAN, and THAILAND.

Quantity

388 units