FDA Recall Terminated

Panda iRes and Giraffe Infant Warmers, labeled in part*** GE Healthcare 8880 Gorman Road Laurel, MD 20723 U.S.A. www.gehealthcare.com*** Infant radiant warmers provide infrared heat in a controlled manner to neonates who are unable to thermo-regulate based on their own physiology.

Recall: Z-2227-2008 · Initiated May 20, 2008

Recall

Recall Number
Z-2227-2008
Event Number
48603
Firm
Datex Ohmeda, Inc. dba GE Healthcare
FEI Number
1121732
Product Code
FMT
Status
Terminated
Root Cause
Software design
Initiated
May 20, 2008
Posted
September 11, 2008
Terminated
May 11, 2012
Address
8880 Gorman Rd, Laurel, MD, 20723-5800

Description

Panda iRes and Giraffe Infant Warmers, labeled in part*** GE Healthcare 8880 Gorman Road Laurel, MD 20723 U.S.A. www.gehealthcare.com*** Infant radiant warmers provide infrared heat in a controlled manner to neonates who are unable to thermo-regulate based on their own physiology.

Reason

Medical device software may be subject to signal interference of the Hands Free Alarm Silence feature that could impact patient safety.

Action

The recalling firm notified Hospital Administrators, NICU Nursing Manager/Labor & Delivery Nursing Managers and Heads of Biomedical Engineering departments at consignee locations by certified letter on 05/20/08. The notification was flagged as "Urgent Medical Device Correction" and advised consignees of safety issues and recommended actions when using the device. The notification advised that the safest way to mitigate the potential interference issue is to disable the Hands Free Alarm Silence. Users were informed that once a permanent correction action is identified, service representatives will implement the correction. For assistance, contact GE Healthcare at 1-800-345-2700.

Distribution

The medical devices were distributed to hospitals nationwide.

Quantity

639 devices