Fresenius Granuflo 1 Mixer Dissolution Unit - Tested, and Granuflo Dissolution Tank, Model Numbers G047-80101 and RTLG047-80101, Manufactured by Fresenius USA Inc., Walnut Creek, CA
Recall
- Recall Number
- Z-2226-2010
- Event Number
- 54546
- Firm
- Fresenius Medical Care North America
- FEI Number
- 3001451489
- Product Code
- KPO
- Status
- Terminated
- Root Cause
- Component design/selection
- Initiated
- November 20, 2009
- Posted
- August 15, 2010
- Terminated
- February 7, 2011
- Address
- 920 Winter St, Waltham, MA, 02451-1521
Description
Fresenius Granuflo 1 Mixer Dissolution Unit - Tested, and Granuflo Dissolution Tank, Model Numbers G047-80101 and RTLG047-80101, Manufactured by Fresenius USA Inc., Walnut Creek, CA
Power cords may fail if the prongs crack or fail at the plug, and may become a fire/burn/shock hazard or delay/interrupt therapy.
An urgent device recall notification was sent to all consignees. Customers are asked to inspect cords for defects, and are refered to preventive maintenance requirements in the manual. Customers who identify the defective cords will be instructed to discontinue use as soon as possible without jeopardizing patient care. Customers are instructed to monitor cords regularly, and pay attention to machines that are frequently plugged or unplugged. Personnel are instructed to not use the cord to unplug the device from the wall (yanking the plug from the wall) and to refrain from rolling over the cord or plug.
Product distributed to 5404 customers throughout the US and the world.