FDA Recall Terminated

GE 1.5T and 3.0T Signa HDX MR System Model Number(s): 2226300, 2377062-61, 2395001, 2395001-2, 5107849, and 5127452. This system is indicated for use as a diagnostic imaging device to produce axial sagittula, coronal and oblique images, spectroscopic images, and or spectra, dynamic images of the internal structure or organs of the entire body,

Recall: Z-2225-2008 · Initiated March 28, 2008

Recall

Recall Number
Z-2225-2008
Event Number
48689
Firm
GE Healthcare
FEI Number
2126677
Product Code
LNH
Status
Terminated
Root Cause
Software change control
Initiated
March 28, 2008
Posted
September 11, 2008
Terminated
December 17, 2011
Address
3000 N Grandview Blvd, Waukesha, WI, 53188-1615

Description

GE 1.5T and 3.0T Signa HDX MR System Model Number(s): 2226300, 2377062-61, 2395001, 2395001-2, 5107849, and 5127452. This system is indicated for use as a diagnostic imaging device to produce axial sagittula, coronal and oblique images, spectroscopic images, and or spectra, dynamic images of the internal structure or organs of the entire body,

Reason

Possible misdiagnosis or inappropriate treatment based on the image. Signal homogeneity may be affected when BRAVO applications are set with IR Prep off. The concern is that signal inhomogeneity can result in an erroneous diagnosis.

Action

Customer Letter and Operator Manual addendum were sent directly to customers in August 2008. When the software correction is issued, GE Field Engineer will visit each site to implement the correction. (Not conducted yet)

Distribution

AK, AL, AR, AZ, CA, CO, CT, DC, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS,MT, NC, ND, NE, NJ, NV, NY, OH, OK, OR, PA, SC, SD, TN, TX, UT, VA, VT, WA,WI, and WV. Countries of Argentina Australia Austria Belgium Brazil Canada Chile China Czech Republic Denmark Finland France French Polynesia Germany Greece Hong Kong India Indonesia Ireland Israel Italy Japan Korea Kuwait Lebanon Malaysia Mexico Netherlands New Zealand Norway Peru Philippines Russia Saudi Arabia Singapore South Africa Spain Sweden Switzerland Taiwan Taiwan Thailand Turkey United Arab Emirates and United Kingdom.

Quantity

720 units (360 US / 360 OUS) estimate