FDA Recall Terminated

Varian brand Clinac and TrueBeam, High Energy Linear Accelerator, Model Numbers: H14, H19, H29; Product is manufactured and distributed by Varian Medical Systems Inc., Palo Alto, CA Indicated for stereotactic radiosurgery and precision radiotherapy for lesions, tumors and conditions anywhere in the body when radiation treatment is indicated.

Recall: Z-2224-2011 · Initiated March 31, 2011

Recall

Recall Number
Z-2224-2011
Event Number
58558
Firm
Varian Medical Systems, Inc. Oncology Systems
FEI Number
2916710
Product Code
IYE
Status
Terminated
Root Cause
Process design
Initiated
March 31, 2011
Posted
May 20, 2011
Terminated
March 13, 2012
Address
911 Hansen Way, Palo Alto, CA, 94304-1028

Description

Varian brand Clinac and TrueBeam, High Energy Linear Accelerator, Model Numbers: H14, H19, H29; Product is manufactured and distributed by Varian Medical Systems Inc., Palo Alto, CA Indicated for stereotactic radiosurgery and precision radiotherapy for lesions, tumors and conditions anywhere in the body when radiation treatment is indicated.

Reason

The coolant may leak. The resulting coolant leak presents a risk of electrical shock to any person working within the protective housing.

Action

On 3/31/2011, all consignees were notified via an Urgent Medical Device Correction letter and on-site visit by the firm's sales representative, informing them of the affected products, the issue, and recommended actions. If non-Varian personnel service the device, then they should have been immediately notified of the content of the letter. Anyone who performs work on the device should exercise extreme care to identify any possible coolant leak and immediately locate and correct the source of the leak. Customers were to immediately notify Varian Service or their internal service personnel of any coolant leakage. Varian was to contact customers to arrange for inspection if one had not been performed. The appropriate personnel should have been advised of the content of the letter. If further clarification was needed, customers were to contact their local Varian Customer Support.

Distribution

Worldwide Distribution

Quantity

718 units