Fresenius 2008H Dialysate Delivery Machine, Manufactured by Fresenius USA Inc., Walnut Creek, CA

Recall: Z-2224-2010 · Initiated November 20, 2009

Recall

Recall Number
Z-2224-2010
Event Number
54546
Firm
Fresenius Medical Care North America
FEI Number
3001451489
Product Code
FKP
Status
Terminated
Root Cause
Component design/selection
Initiated
November 20, 2009
Posted
August 15, 2010
Terminated
February 7, 2011
Address
920 Winter St, Waltham, MA, 02451-1521

Description

Fresenius 2008H Dialysate Delivery Machine, Manufactured by Fresenius USA Inc., Walnut Creek, CA

Reason

Power cords may fail if the prongs crack or fail at the plug, and may become a fire/burn/shock hazard or delay/interrupt therapy.

Action

An urgent device recall notification was sent to all consignees. Customers are asked to inspect cords for defects, and are refered to preventive maintenance requirements in the manual. Customers who identify the defective cords will be instructed to discontinue use as soon as possible without jeopardizing patient care. Customers are instructed to monitor cords regularly, and pay attention to machines that are frequently plugged or unplugged. Personnel are instructed to not use the cord to unplug the device from the wall (yanking the plug from the wall) and to refrain from rolling over the cord or plug.

Distribution

Product distributed to 5404 customers throughout the US and the world.

Quantity

33