FDA Recall
Terminated
IDEAL IMPLANT Structured Breast Implant, saline breast implant, in the following sizes: 210cc, 240cc, 270cc, 300cc, 335cc, 370cc, 405cc, 440cc, 475cc, 515cc, 555cc, 595cc, 635cc, and 675cc.
Recall: Z-2219-2019
·
Initiated March 18, 2019
Recall
- Recall Number
- Z-2219-2019
- Event Number
- 83251
- Firm
- Ideal Implant Incorporated
- FEI Number
- 3008342021
- Product Code
- FWM
- Status
- Terminated
- Root Cause
- Process control
- Initiated
- March 18, 2019
- Terminated
- March 3, 2021
- Address
- 14881 Quorum Dr, Ste 925, Dallas, TX, 75254-7132
Description
IDEAL IMPLANT Structured Breast Implant, saline breast implant, in the following sizes: 210cc, 240cc, 270cc, 300cc, 335cc, 370cc, 405cc, 440cc, 475cc, 515cc, 555cc, 595cc, 635cc, and 675cc.
Reason
The implants were deflating due to a piece of silicone disrupting the seal of the posterior valve shortly after implantation.
Action
The recall was initiated via telephone and email on 3/18/2019 to one consignee and via email on 3/22/2019 to the other consignees and requested return of the implants.
Distribution
Worldwide distribution - US Nationwide distribution in the states of AL, AR, AZ, CA, CO, CT, DC, FL, GA, HI, IA, IL, IN, KS, KY, LA, MA, MD, MI, MO, NC, ND, NH, NJ, NY, OH, OK, OR, PA, SC, SD, TN, TX, UT, VA, WA, WI, and WV. Country of Canada.
Quantity
1,704 implants