Hi-Art System, H-0000-0003 Usage: The TomoTherapy Hi-Art System is intended to be used as an integrated system for the planning and precise delivery of radiation therapy, stereotactic radiotherapy, or stereotactic radiosurgery to tumors or other targeted tissues while minimizing the delivery of radiation to vital healthy tissue. The megavoltage x-ray radiation is delivered in a rotational, non-rotational, modulated (IMRT), or non modulated (non-IMRT/three dimensional conformal) format in accordance with the physician approved plan

Recall: Z-2219-2010 · Initiated December 1, 2009

Recall

Recall Number
Z-2219-2010
Event Number
56166
Firm
TomoTherapy Incorporated
FEI Number
3003873069
Product Code
MUJ
Status
Terminated
Root Cause
Software design
Initiated
December 1, 2009
Posted
August 12, 2010
Terminated
December 17, 2011
Address
1240 Deming Way, Madison, WI, 53717

Description

Hi-Art System, H-0000-0003 Usage: The TomoTherapy Hi-Art System is intended to be used as an integrated system for the planning and precise delivery of radiation therapy, stereotactic radiotherapy, or stereotactic radiosurgery to tumors or other targeted tissues while minimizing the delivery of radiation to vital healthy tissue. The megavoltage x-ray radiation is delivered in a rotational, non-rotational, modulated (IMRT), or non modulated (non-IMRT/three dimensional conformal) format in accordance with the physician approved plan

Reason

An issue was identified with the TomoTherapy HI-Art System. In the event a patient or DQA plan has a moved image, roll adjustments applied during registration will be incorrect. The Planning Station Plan Settings and DQA Setup tabs allow for images to be moved during planning. During registration when roll is applied on moved images, the Operator Station incorrectly rolls the image about

Action

Consignee was sent a TomoTherapy "Urgent Medical device Correction Field Safety Notice' dated December 01, 2009. The letter described the Issue, Product Affected, Recommended Action and Resolution.

Distribution

VA