OneLIF Interbody Cage (REF/Description): 010-107-1007/Small 7 Degree 10mm X 7mm, 010-107-1007-2/Small 7 Degree 10mm X 7mm - Oblique, 010-107-1209/Small 7 Degree 12mm X 9mm, 010-107-1411/Small 7 Degree 14mm X 11mm, 010-107-1613/Small 7 Degree 16mm X 13mm, 010-113-1006/Small, 13 Degree 10mm X 6mm, 010-113-1006-2/Small, 13 Degree, 10mm X 6mm - Oblique, 010-113-1208/Small 13 Degree 12mm X 8mm, 010-113-1410/Small 13 Degree 14mm X 10mm, 010-113-1612/Small 13 Degree 16mm X 12mm, 010-119-1407/19 Degree 14mm X 7mm, 010-119-1609/Small 19 Degree 16mm X 9mm, 010-119-1811/Small 19 Degree 18mm X 11mm, 010-125-1606/Small 25 Degree 16mm X 6mm, 010-125-1808/Small 25 Degree 18mm X 8mm, 010-207-1007/Medium 7 Degree 10mm X 7mm, 010-207-1007-2/Medium 7 Degree 10mm X 7mm - Oblique, 010-207-1209/Medium 7 Degree 12mm X 9mm, 010-207-1411/Medium 7 Degree 14mm X 11mm, 010-207-1613/Medium 7 Degree 16mm X 13mm, 010-213-1206/Medium 13 Degree 12mm X 6mm, 010-213-1408/Medium 13 Degree 14mm X 8mm, 010-213-1610/Medium 13 Degree 16mm X 10mm, 010-213-1812/Medium 13 Degree 18mm X 12mm, 010-219-1406/Medium 19 Degree 14mm X 6mm, 010-219-1608/Medium 19 Degree 16mm X 8mm, 010-219-1810/Medium 19 Degree 18mm X 10mm, 010-225-1807/Medium 25 Degree 18mm X 7mm, 010-225-2009/Medium 25 Degree 20mm X 9mm, 010-307-1006/Large 7 Degree,10mm X 6mm, 010-307-1006-2/Large 7 Degree 10mm X 6mm - Oblique, 010-307-1208/ Large 7 Degree 12mm X 8mm, 010-307-1410/Large 7 Degree 14mm X 10mm, 010-307-1612/Large 7 Degree 16mm X 12mm, 010-313-1205/Large 13 Degree 12mm X 5mm, 010-313-1407/Large 13 Degree 14mm X 7mm, 010-313-1609/Large 13 Degree 16mm X 9mm, 010-313-1811/Large 13 Degree 18mm X 11mm, 010-313-2013/Large 13 Degree 20mm X 13mm, 010-319-1606/Large 19 Degree 16mm X 6mm, 010-319-1808/ Large 19 Degree 18mm X 8mm, 010-319-2010/ Large 19 Degree 20mm X 10mm, 010-325-2007/Large 25 Degree 20mm X 7mm, 010-325-2209/ Large 25 Degree 22mm X 9mm, Implant Kits containing OneLIF Interbody Cages, Name/Kit Family: Kit-OneLIF-10 mm (716)/Kit-OneLIF-10 mm, Kit-OneLIF-10 mm (717)/Kit-OneLIF-10 mm, Kit-OneLIF-10 mm (718)/Kit-OneLIF-10 mm, Kit-OneLIF-10 mm (719)/Kit-OneLIF-10 mm, Kit-OneLIF-10 mm (722)/Kit-OneLIF-10 mm, Kit-OneLIF-10 mm (723)/Kit-OneLIF-10 mm, Kit-OneLIF-10 mm (724)/Kit-OneLIF-10 mm, Kit-OneLIF-10 mm (733)/Kit-OneLIF-10 mm, Kit-OneLIF-10 mm (734)/Kit-OneLIF-10 mm, Kit-OneLIF-10 mm (737)/Kit-OneLIF-10 mm, Kit-OneLIF-10 mm (738)/Kit-OneLIF-10 mm, Kit-OneLIF-10 mm (746)/Kit-OneLIF-10 mm, Kit-OneLIF-10 mm (751)/Kit-OneLIF-10 mm, Kit-OneLIF-10 mm (761)/Kit-OneLIF-10 mm, Kit-OneLIF-10 mm (762)/Kit-OneLIF-10 mm, Kit-OneLIF-10 mm (787)/Kit-OneLIF-10 mm, Kit-OneLIF- IMPOneLIF Implant Kit (701)/Kit-OneLIF-IMP - OneLIF Implant Kit, Kit-OneLIF-IMP - OneLIF Implant Kit (702)/Kit-OneLIF-IMP - OneLIF Implant Kit, Kit-OneLIF-IMP - OneLIF Implant Kit (703) NL/Kit-OneLIF-IMP - OneLIF Implant Kit, Kit-OneLIF-IMP - OneLIF Implant Kit (704)/Kit-OneLIF-IMP - OneLIF Implant Kit, Kit-OneLIF-IMP - OneLIF Implant Kit (705)/Kit-OneLIF-IMP - OneLIF Implant Kit, Kit-OneLIF-IMP - OneLIF Implant Kit (706)/Kit-OneLIF-IMP - OneLIF Implant Kit, Kit-OneLIF-IMP - OneLIF Implant Kit (708)/Kit-OneLIF-IMP - OneLIF Implant Kit, Kit-OneLIF-IMP - OneLIF Implant Kit (709)/Kit-OneLIF-IMP - OneLIF Implant Kit,
Recall
- Recall Number
- Z-2213-2026
- Event Number
- 98835
- Firm
- Novapproach Spine, LLC
- FEI Number
- 3020307303
- Product Code
- OVD
- Status
- Open, Classified
- Root Cause
- Process design
- Initiated
- April 23, 2026
- Posted
- May 20, 2026
- Address
- 13900 Tech City Cir, Ste 300, Alachua, FL, 32615-6091
Description
OneLIF Interbody Cage (REF/Description): 010-107-1007/Small 7 Degree 10mm X 7mm, 010-107-1007-2/Small 7 Degree 10mm X 7mm - Oblique, 010-107-1209/Small 7 Degree 12mm X 9mm, 010-107-1411/Small 7 Degree 14mm X 11mm, 010-107-1613/Small 7 Degree 16mm X 13mm, 010-113-1006/Small, 13 Degree 10mm X 6mm, 010-113-1006-2/Small, 13 Degree, 10mm X 6mm - Oblique, 010-113-1208/Small 13 Degree 12mm X 8mm, 010-113-1410/Small 13 Degree 14mm X 10mm, 010-113-1612/Small 13 Degree 16mm X 12mm, 010-119-1407/19 Degree 14mm X 7mm, 010-119-1609/Small 19 Degree 16mm X 9mm, 010-119-1811/Small 19 Degree 18mm X 11mm, 010-125-1606/Small 25 Degree 16mm X 6mm, 010-125-1808/Small 25 Degree 18mm X 8mm, 010-207-1007/Medium 7 Degree 10mm X 7mm, 010-207-1007-2/Medium 7 Degree 10mm X 7mm - Oblique, 010-207-1209/Medium 7 Degree 12mm X 9mm, 010-207-1411/Medium 7 Degree 14mm X 11mm, 010-207-1613/Medium 7 Degree 16mm X 13mm, 010-213-1206/Medium 13 Degree 12mm X 6mm, 010-213-1408/Medium 13 Degree 14mm X 8mm, 010-213-1610/Medium 13 Degree 16mm X 10mm, 010-213-1812/Medium 13 Degree 18mm X 12mm, 010-219-1406/Medium 19 Degree 14mm X 6mm, 010-219-1608/Medium 19 Degree 16mm X 8mm, 010-219-1810/Medium 19 Degree 18mm X 10mm, 010-225-1807/Medium 25 Degree 18mm X 7mm, 010-225-2009/Medium 25 Degree 20mm X 9mm, 010-307-1006/Large 7 Degree,10mm X 6mm, 010-307-1006-2/Large 7 Degree 10mm X 6mm - Oblique, 010-307-1208/ Large 7 Degree 12mm X 8mm, 010-307-1410/Large 7 Degree 14mm X 10mm, 010-307-1612/Large 7 Degree 16mm X 12mm, 010-313-1205/Large 13 Degree 12mm X 5mm, 010-313-1407/Large 13 Degree 14mm X 7mm, 010-313-1609/Large 13 Degree 16mm X 9mm, 010-313-1811/Large 13 Degree 18mm X 11mm, 010-313-2013/Large 13 Degree 20mm X 13mm, 010-319-1606/Large 19 Degree 16mm X 6mm, 010-319-1808/ Large 19 Degree 18mm X 8mm, 010-319-2010/ Large 19 Degree 20mm X 10mm, 010-325-2007/Large 25 Degree 20mm X 7mm, 010-325-2209/ Large 25 Degree 22mm X 9mm, Implant Kits containing OneLIF Interbody Cages, Name/Kit Family: Kit-OneLIF-10 mm (716)/Kit-OneLIF-10 mm, Kit-OneLIF-10 mm (717)/Kit-OneLIF-10 mm, Kit-OneLIF-10 mm (718)/Kit-OneLIF-10 mm, Kit-OneLIF-10 mm (719)/Kit-OneLIF-10 mm, Kit-OneLIF-10 mm (722)/Kit-OneLIF-10 mm, Kit-OneLIF-10 mm (723)/Kit-OneLIF-10 mm, Kit-OneLIF-10 mm (724)/Kit-OneLIF-10 mm, Kit-OneLIF-10 mm (733)/Kit-OneLIF-10 mm, Kit-OneLIF-10 mm (734)/Kit-OneLIF-10 mm, Kit-OneLIF-10 mm (737)/Kit-OneLIF-10 mm, Kit-OneLIF-10 mm (738)/Kit-OneLIF-10 mm, Kit-OneLIF-10 mm (746)/Kit-OneLIF-10 mm, Kit-OneLIF-10 mm (751)/Kit-OneLIF-10 mm, Kit-OneLIF-10 mm (761)/Kit-OneLIF-10 mm, Kit-OneLIF-10 mm (762)/Kit-OneLIF-10 mm, Kit-OneLIF-10 mm (787)/Kit-OneLIF-10 mm, Kit-OneLIF- IMPOneLIF Implant Kit (701)/Kit-OneLIF-IMP - OneLIF Implant Kit, Kit-OneLIF-IMP - OneLIF Implant Kit (702)/Kit-OneLIF-IMP - OneLIF Implant Kit, Kit-OneLIF-IMP - OneLIF Implant Kit (703) NL/Kit-OneLIF-IMP - OneLIF Implant Kit, Kit-OneLIF-IMP - OneLIF Implant Kit (704)/Kit-OneLIF-IMP - OneLIF Implant Kit, Kit-OneLIF-IMP - OneLIF Implant Kit (705)/Kit-OneLIF-IMP - OneLIF Implant Kit, Kit-OneLIF-IMP - OneLIF Implant Kit (706)/Kit-OneLIF-IMP - OneLIF Implant Kit, Kit-OneLIF-IMP - OneLIF Implant Kit (708)/Kit-OneLIF-IMP - OneLIF Implant Kit, Kit-OneLIF-IMP - OneLIF Implant Kit (709)/Kit-OneLIF-IMP - OneLIF Implant Kit,
Intervertebral body fusion system straight inserter failed to properly attach to affected cages, which may result in an inability to securely engage the implant with the inserter, which could lead to procedural delay and/or the need to select an alternative implant.
On 4/23/2026, removal notices were sent to customers who were asked to do the following: 1) Return affected devices using the return label. 2) Firm is providing unaffected kits. If you have recall-related questions contact firm at [email protected]
US Nationwide distribution in the states of NY, CT, CA, TX, FL, MI.