Medical Positioning Rapid Response HUT table, Powered Medical Examination table, Model # 1219. Product Usage: Used to conduct syncope examinations.
Recall
- Recall Number
- Z-2211-2014
- Event Number
- 68763
- Firm
- Medical Positioning Inc
- FEI Number
- 1000118382
- Product Code
- LGX
- Status
- Terminated
- Root Cause
- Component design/selection
- Initiated
- July 10, 2014
- Posted
- August 13, 2014
- Terminated
- January 3, 2017
- Address
- 1717 Washington St, Kansas City, MO, 64108-1121
Description
Medical Positioning Rapid Response HUT table, Powered Medical Examination table, Model # 1219. Product Usage: Used to conduct syncope examinations.
Excessive weight on the table will cause tables with the affected component to drop into the Trendelenburg position more rapidly than normal. The product was manufactured with a component that can cause the device to malfunction.
Medical Positioning Inc. sent Potential Product Hazard Bulletin letter dated July 8, 2014 to affected customers, via USPS certified return receipt mail. The letter identified the affected product, problem and actions to be taken. Customers were informed a Product Support Specialist will contact them to make arrangements for a Field Service Technician to perform a test designed to identify if the table component affects the table mechanics. If the table is identified as having the affected component, the component will be replaced.
Worldwide Distribution - US Nationwide in the states of CA, GA, FL, IL, IN, KS, KY, MD, MN, MO, MS, NC, NY, OH, TN, TX, VA and WI.