Preamendments device-Sarns Dual Cooler-Heater. The Sarns Dual Cooler-Heater is a self-contained unit for supplying cold and warm water to heat exchangers during open heart surgery. It supplies water at a rate of up to 6.5 gallons per minute (25 liters per minute) at zero head, and will provide a maximum pressure output of 11 pounds per square inch (76,000 pascals). The Sarns TM Dual Cooler-Heater can be used with heat exchangers capable of accepting these flows and pressures. Product Usage: The Sarns Dual Cooler-Heater is indicated for use as a water supply system to a blood heat exchanger in open-heart surgery.
Recall
- Recall Number
- Z-2210-2016
- Event Number
- 74389
- Firm
- Terumo Cardiovascular Systems Corporation
- FEI Number
- 1828100
- Product Code
- DWC
- Status
- Terminated
- Root Cause
- Nonconforming Material/Component
- Initiated
- June 8, 2016
- Posted
- July 14, 2016
- Terminated
- August 4, 2017
- Address
- 6200 Jackson Rd, Ann Arbor, MI, 48103
Description
Preamendments device-Sarns Dual Cooler-Heater. The Sarns Dual Cooler-Heater is a self-contained unit for supplying cold and warm water to heat exchangers during open heart surgery. It supplies water at a rate of up to 6.5 gallons per minute (25 liters per minute) at zero head, and will provide a maximum pressure output of 11 pounds per square inch (76,000 pascals). The Sarns TM Dual Cooler-Heater can be used with heat exchangers capable of accepting these flows and pressures. Product Usage: The Sarns Dual Cooler-Heater is indicated for use as a water supply system to a blood heat exchanger in open-heart surgery.
Terumo Cardiovascular Systems (Terumo CVS) is conducting a voluntary recall of its heater-cooler devices due to the distributor no longer selling the Inter chlor sanitizer (Product Code 205) referenced in its heater-cooler cleaning instructions.
Terumo Cardiovascular Systems (Terumo CVS) initiated a voluntary recall on 06/08/2016, by mailing certifed letters to customers. Terumo CVS is notifying all users of its heater-cooler devices that the distributor, Intercon Chemical Company, no longer sells the Interchlor sanitizer (Product Code 205) referenced in its heater-cooler cleaning instructions. However, the identical sanitizer is available from other sources. This notice applies to the following Terumo CVS heater-cooler devices: Sarns" TCM and TCM II Cooling and Heating Systems Sarns" Dual Cooler-Heaters HX2" Temperature Management Systems Terumo CVS will provide users with updated user instructions regarding the 10% Sodium Hypochlorite sanitization solution. Instructions will be provided under separate cover, when available; updated manual and cleaning guide will be completed in December 2016. Customers are asked to do the following: Review this notice and ensure that all users receive notice of this issue. 2. Return the attached Customer Response Form acknowledging receipt of this notice. 3. Source a replacement for the Interchlor sanitizer. See information above or call Terumo CVS Technical Support for assistance. If you have questions about this notification, please contact your Terumo representative or Terumo CVS Technical Support: 800.441.3220, extension 6932 Technical Support Hours: Monday Friday 8 a.m. 6 p.m. ET (24/7 Emergency Hours)
Domestic: AK, AL, AR, AZ, CA , CO, CT, DC , DE , FL, GA, HI, IA, ID, IL, IN, KS, KY,LA , MA, MD, ME, MI , MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV , NY, OH, OK, OR, PA, PR, RI, SC, SD, TN, TX, UT, VA, WA, WI, WV, WY Foreign: Argentina, Australia, Belgium,Brazil, Canada, Chile, China, Colombia, Costa Rica, Dominican Republic, Ecuador, El Salvador, England, Finland, France, Germany Greece, Hong Kong, India, Indonesia, Iran, Ireland, Israel, Italy, Jamaica, Japan, Jordan, Korea, Lebanon, Malaysia Malta, Mexico, Netherlands, New Zealand, Norway, Philippines, Portugal, Russia, Saudi Arabia, Singapore, South Korea, Spain, Sweden, Switzerland, Taiwan, Thailand, Turkey, United Arab Emirates , Uruguay , Venezuela, Vietnam
6412 units