Brilliance CT 6, M/N 728256; Brilliance CT 16 M/N 728246; Brilliance CT 40, M/N 728235 & Brilliance CT 64 M/N 728231, Philips Healthcare, Cleveland, OH 44143. Intended to produce cross-sectional images of the body by computer reconstruction.
Recall
- Recall Number
- Z-2210-2013
- Event Number
- 65691
- Firm
- Philips Medical Systems (Cleveland) Inc
- FEI Number
- 1525965
- Product Code
- JAK
- Status
- Terminated
- Root Cause
- Component design/selection
- Initiated
- July 5, 2013
- Posted
- September 16, 2013
- Terminated
- September 15, 2015
- Address
- 595 Miner Rd, Cleveland, OH, 44143-2131
Description
Brilliance CT 6, M/N 728256; Brilliance CT 16 M/N 728246; Brilliance CT 40, M/N 728235 & Brilliance CT 64 M/N 728231, Philips Healthcare, Cleveland, OH 44143. Intended to produce cross-sectional images of the body by computer reconstruction.
The patient support may move in an unintended manner if the footswitch cover impinges on the footswitch.
Philips Healthcare sent an "URGENT-MEDICAL DEVICE CORRECTION" letter dated June 4, 2013 to all affected customers. The letter identified the product, problem, and actions to be taken by the customers. Contact the Customer Care Center at 1-800-722-9377, option 5 for information concerning this issue.
Worldwide Distribution-USA (nationwide) including DC and the states of AL, AR, AZ, CA, CO, DE, FL, GA, IN, KY, MA, MD, MI, MN, MS, NC, NM, NV, NY, OH, OK, PA, SC, TN, TX, VT, WI & WV, and the countries of Australia, Austria, Bosnia & Herzegovina, Chile, China, Croatia, Denmark, Equatorial Guinea, France, Germany, Greenland, Iceland, Indonesia, Israel, Japan, Kazakhstan, Netherlands, Nigeria, Norway, Poland, Romania, South Africa, South Korea, Sweden, Switzerland, Taiwan, United Arab Emirates, United Kingdom & Vietnam.
M/N: 728256, 4 units; M/N: 728246, 89 units; M/N: 728235, 9 units & M/N: 728231, 207 units