FDA Recall Terminated

Medical Positioning Rapid Response HUT table, Powered Medical Examination table, Model # 1019. Product Usage: Used to conduct syncope examinations.

Recall: Z-2209-2014 · Initiated July 10, 2014

Recall

Recall Number
Z-2209-2014
Event Number
68763
Firm
Medical Positioning Inc
FEI Number
1000118382
Product Code
LGX
Status
Terminated
Root Cause
Component design/selection
Initiated
July 10, 2014
Posted
August 13, 2014
Terminated
January 3, 2017
Address
1717 Washington St, Kansas City, MO, 64108-1121

Description

Medical Positioning Rapid Response HUT table, Powered Medical Examination table, Model # 1019. Product Usage: Used to conduct syncope examinations.

Reason

Excessive weight on the table will cause tables with the affected component to drop into the Trendelenburg position more rapidly than normal. The product was manufactured with a component that can cause the device to malfunction.

Action

Medical Positioning Inc. sent Potential Product Hazard Bulletin letter dated July 8, 2014 to affected customers, via USPS certified return receipt mail. The letter identified the affected product, problem and actions to be taken. Customers were informed a Product Support Specialist will contact them to make arrangements for a Field Service Technician to perform a test designed to identify if the table component affects the table mechanics. If the table is identified as having the affected component, the component will be replaced.

Distribution

Worldwide Distribution - US Nationwide in the states of CA, GA, FL, IL, IN, KS, KY, MD, MN, MO, MS, NC, NY, OH, TN, TX, VA and WI.

Quantity

2 beds