FDA Recall Terminated

COBAS AmpliPrep Sample Tube Input Barcode Clips, catalog number 03137040001. Input S-tube barcode clips hold sample input tubes in the COBAS AmpliPrep Instrument (CAP) sample racks and have a barcode to identify the sample.

Recall: Z-2209-2012 · Initiated February 1, 2012

Recall

Recall Number
Z-2209-2012
Event Number
62586
Firm
Roche Molecular Systems, Inc.
FEI Number
2243471
Product Code
JJF
Status
Terminated
Root Cause
Nonconforming Material/Component
Initiated
February 1, 2012
Posted
August 15, 2012
Terminated
July 16, 2013
Address
1080 Us Highway 202 S, Branchburg, NJ, 08876-3733

Description

COBAS AmpliPrep Sample Tube Input Barcode Clips, catalog number 03137040001. Input S-tube barcode clips hold sample input tubes in the COBAS AmpliPrep Instrument (CAP) sample racks and have a barcode to identify the sample.

Reason

Certain lots of COBAS AmpliPrep (CAP) S-tube Input (Box of 12 bags 12x24 S-tube and 12x24 S-clip) can cause the sample input tubes to stick in the barcode clips. As a consequence, the CAP instrument will lift both the sample input tube together with the barcode clip from the sample rack, resulting in an aborted run. No erroneous results would be produced if the issue occurs, but samples in proce

Action

On 1/17/2012 an "Analyzer Bulletin" notification was sent to Roche affiliates and on 2/1/2012 an "Analyzer Bulletin" notification was sent to all affected US Consignees. The notification described the product issue and provided recommended actions. Customers were referred to Roche Technical Support at 800-526-1247 for questions.

Distribution

Nationwide Distribution (USA).

Quantity

2,132 boxes