FDA Recall Terminated

Brilliance Big Bore Computed Tomography X-Ray System, Model Number: 728243. The units subject to correction are equipped with software version: 3.5.17030. Intended to produce cross-sectional images of the body by computer reconstruction of x-ray transmission data taken at different angles and planes.

Recall: Z-2207-2011 · Initiated March 7, 2011

Recall

Recall Number
Z-2207-2011
Event Number
58510
Firm
Philips Medical Systems (Cleveland) Inc
FEI Number
1525965
Product Code
JAK
Status
Terminated
Root Cause
Other
Initiated
March 7, 2011
Posted
May 19, 2011
Terminated
February 22, 2013
Address
595 Miner Road, Cleveland, OH, 44143-2131

Description

Brilliance Big Bore Computed Tomography X-Ray System, Model Number: 728243. The units subject to correction are equipped with software version: 3.5.17030. Intended to produce cross-sectional images of the body by computer reconstruction of x-ray transmission data taken at different angles and planes.

Reason

A program software error in the algorithm produces image artifacts that could possibly lead to a misdiagnosis of patients who have small metal implants.

Action

A 'Field Safety Notice - Urgent - Medical Device Correction' notification was sent to the firm's customers via Certified mail on 3/8/2011. The notification letter explains the nature of the Metal Artifact Reduction/Metal Artifact Correction software error; the Brilliance CT X-Ray systems subject to correction and how to identify them; and the circumstances under which this error can occur. It also asks that the user stop running the MAR application on any of the affected Brilliance systems identified with the recalled serial numbers and software versions alluded to in the medical device correction notification. Lastly, the notification states that Philips Medical will provide a solution that will fix the issue and the notification lists contact names; telephone numbers; and E-mail addresses which the customer can contact for further information concerning this device issue, or customer support assistance.

Distribution

Nationwide Distribution -- DC, IL, IN, MI, MO, NY, OH, OR, PA, TX, and VT.

Quantity

5 units