AGFA Digital Radiography X- Ray system DR 800 with MUSICA Dynamic, Type number: 6010/200 - Product Usage: DR 800: Agfa s DR 800 system is indicated for performing dynamic imaging examinations (fluoroscopy and/or rapid sequence) of the following anatomies/procedures: positioning fluoroscopy procedures, gastro-intestinal examinations, urogenital tract examinations, and angiography. It is intended to replace fluoroscopic images obtained through image intensifier technology. In addition, the system is intended for project radiography of all body parts. The DR 800 is not intended for mammography applications

Recall: Z-2202-2020 · Initiated April 6, 2020

Recall

Recall Number
Z-2202-2020
Event Number
85599
Firm
Agfa N.V. Septestraat 27 Mortsel Belgium
FEI Number
3001556265
Product Code
JAA
Status
Terminated
Root Cause
Software design
Initiated
April 6, 2020
Terminated
April 3, 2023

Description

AGFA Digital Radiography X- Ray system DR 800 with MUSICA Dynamic, Type number: 6010/200 - Product Usage: DR 800: Agfa s DR 800 system is indicated for performing dynamic imaging examinations (fluoroscopy and/or rapid sequence) of the following anatomies/procedures: positioning fluoroscopy procedures, gastro-intestinal examinations, urogenital tract examinations, and angiography. It is intended to replace fluoroscopic images obtained through image intensifier technology. In addition, the system is intended for project radiography of all body parts. The DR 800 is not intended for mammography applications

Reason

Under specific conditions (Fluoroscopic exam, ABS=OFF, manual change of parameters) wrong calculation of the dose/minute for fluoroscopy exams can be possible

Action

Agfa issued Urgent Field Safety Notice via email on 4/21/20 to consignees, letter states reason for recall, health risk and action to take: Acknowledgment, via a feedback form or email, that the information was received and understood was requested from the consignees.

Distribution

US Nationwide distribution including in the sates of FL, IL, IN, LA, MA, MO, OH, SC, VA, WI.

Quantity

13 units