FDA Recall Terminated

Beta-Hydroxybutyrate Reagent A - Enzyme (unlabeled). in vitro diagnostic reagent.

Recall: Z-2200-2021 · Initiated July 12, 2021

Recall

Recall Number
Z-2200-2021
Event Number
88323
FEI Number
1616487
Product Code
JIN
Status
Terminated
Root Cause
Under Investigation by firm
Initiated
July 12, 2021
Terminated
March 29, 2022
Address
1261 N Main St, Stanbio Laboratory, LP, Boerne, TX, 78006-3014

Description

Beta-Hydroxybutyrate Reagent A - Enzyme (unlabeled). in vitro diagnostic reagent.

Reason

Deterioration in the stability of the reagents which has resulted in lower than expected or out of range Quality Control values.

Action

On 07/12/2021, the firm sent an "PRODUCT CORRECTION NOTIFICATION" Letter via First Class mail (U.S.) and email (O.U.S.) to inform customers that the Recalling Firm has observed deterioration in the stability of the liquid reagents which has resulted in lower than expected or out of range Quality Control values. Customers are instructed to: 1) Discontinue use of and quarantine any affected product. 2) Only the lot listed in the table provided are affected and customers may continue to use any other lots they may have in their possession. 3) Share this information with their laboratory staff and retain this notification as part of their laboratory Quality System documentation. 4) If customers have forwarded the product to another laboratory, please provide a copy of this letter to them. Provide the name, address and a contact for the laboratory on the Verification Form. 5) Complete and FAX or email the enclosed Verification Form within 10 days to confirm their receipt of this notice. If you have further questions, please contact Technical Support at 1-800-531-5535 ext. 128.

Distribution

Worldwide distribution - U.S. Nationwide distribution including in the states of AK, AL, AR, AZ, CA, CO, CT, DE, FL, GA, HI, IA, ID, IL, IN, KY, LA, MA, MD, MI, MO, NC, NH, NJ, NV, NY, OH, OK, OR, PA, Puerto Rico, RI, TN, TX, VA, VT, WA, WI, and WV. The countries of Australia, Canada, Chile, Federal Republic of Germany, Hong Kong, Singapore, Taiwan, Thailand, United Kingdom, and Viet Nam.

Quantity

200 units