FDA Recall Terminated

T8 REMOVAL HEAD, REF 8910641, QTY: 1 EA, Product Usage: The non-sterile T8 Removal Diriver is used with the Spire and Spire Z Spinal System in conjunction with the set screwdriver to remove the implant during removal surgery. The set screw driver is attached to the end of the removal head and used to disengage the remaining portion of the set screw within the device.

Recall: Z-2200-2013 · Initiated August 12, 2013

Recall

Recall Number
Z-2200-2013
Event Number
66010
Firm
Medtronic Sofamor Danek USA Inc
FEI Number
1000477302
Product Code
HXX
Status
Terminated
Root Cause
Nonconforming Material/Component
Initiated
August 12, 2013
Posted
September 14, 2013
Terminated
December 27, 2013
Address
1800 Pyramid Place, Memphis, TN, 38132

Description

T8 REMOVAL HEAD, REF 8910641, QTY: 1 EA, Product Usage: The non-sterile T8 Removal Diriver is used with the Spire and Spire Z Spinal System in conjunction with the set screwdriver to remove the implant during removal surgery. The set screw driver is attached to the end of the removal head and used to disengage the remaining portion of the set screw within the device.

Reason

One lot of the T8 Removal Head had incorrect hardness specification, which could result in the T8 Removal head failing during removal of the CD Horizon Spire and Spire Z Spinal System implant. There is a potential for tissue damage to occur and/or minor surgical delay to remove the remaining fragments of the tip.

Action

Medtronic sent an Urgent Voluntary Market Removal Recall letter dated August 9, 2013 with Recall Questionnaires and Confirmation forms included to Risk Managers and Sales Representatives. The letter identified the affected product, problem and action to be taken. Customers were instructed to remove the affected product, discontinue use and contact your Medtronic Sales Representative to coordinate the return and replenishment of affected product. For questions call your Medtronic Sales Representative or Global Quality Department at 800-876-3133.

Distribution

USA Nationwide Distribution in the state of GA, AK, WA, and CA.

Quantity

9 units