Precision Sampling Set, REF: C20206 for use with the CoSense End-Tidal Carbon Monoxide Monitor

Recall: Z-2198-2025 · Initiated April 21, 2025

Recall

Recall Number
Z-2198-2025
Event Number
97184
Firm
Capnia Inc
FEI Number
3008614998
Product Code
CCJ
Status
Open, Classified
Root Cause
Device Design
Initiated
April 21, 2025
Posted
July 29, 2025
Address
1101 Chess Dr, Foster City, CA, 94404-1102

Description

Precision Sampling Set, REF: C20206 for use with the CoSense End-Tidal Carbon Monoxide Monitor

Reason

Sampling set nasal cannula used with End-Tidal Carbon Monoxide (ETCO) monitor may have residual carbon monoxide and volatile organic compounds present due to insufficient aeration of the cannula's carbon filter prior to sealing, which may lead to higher than expected ETOCc scores during clinical evaluations.

Action

Starting on 4/21/2025 Product Advisory and Voluntary Field Action emails were sent to customers who were told affected devices would be replaced. Customers with questions could contact the firm at 904-588-2986

Distribution

US Nationwide distribution in the states of PA, UT, TX, FL, MA, IA.

Quantity

2160