IMMAGE Immunochemistry Systems Buffer 1 (BUF1), Part Number: 447650
Recall
- Recall Number
- Z-2195-2010
- Event Number
- 55787
- Firm
- Beckman Coulter Inc.
- FEI Number
- 2050012
- Product Code
- JQX
- Status
- Terminated
- Root Cause
- Nonconforming Material/Component
- Initiated
- May 21, 2010
- Posted
- August 10, 2010
- Terminated
- May 22, 2012
- Address
- 200 S Kraemer Blvd, Brea, CA, 92821-6208
Description
IMMAGE Immunochemistry Systems Buffer 1 (BUF1), Part Number: 447650
Beckman Coulter is initiating a recall of IMMAGE Immunochemistry Systems Buffer 1 (BUF1) due to received reports of increased recovery of control and patient samples using the lots of IMMAGE BUF1. Some lots of BUF1 have been reported to cause high shifts in recovery of control or patient samples. Quality Control failures may result after replacing BUF1 with a different lot.
Beckman Coulter will send a Product Corrective Action (PCA) letter on the week of April 19, 2010 to the affected customers. The letter provided the customers with an explanation of the problem and a work around.
Worldwide: U.S., Canada, Australia, Austria, Belgium, China, Czech Republic, Denmark, France, Germany, Greece, Hong Kong, Iceland, Ireland, Italy, Japan, Liechtenstein, Macao, Malaysia, Netherlands, New Zealand, Norway, Oman, Poland, Qatar, Russian Federation, Singapore, Slovakia, South Africa, Spain, Sweden, Switzerland, Taiwan, Tunisia, Turkey and United Kingdom,
1309