FDA Recall Terminated

IMMAGE Immunochemistry Systems Buffer 1 (BUF1), Part Number: 447650

Recall: Z-2195-2010 · Initiated May 21, 2010

Recall

Recall Number
Z-2195-2010
Event Number
55787
Firm
Beckman Coulter Inc.
FEI Number
2050012
Product Code
JQX
Status
Terminated
Root Cause
Nonconforming Material/Component
Initiated
May 21, 2010
Posted
August 10, 2010
Terminated
May 22, 2012
Address
200 S Kraemer Blvd, Brea, CA, 92821-6208

Description

IMMAGE Immunochemistry Systems Buffer 1 (BUF1), Part Number: 447650

Reason

Beckman Coulter is initiating a recall of IMMAGE Immunochemistry Systems Buffer 1 (BUF1) due to received reports of increased recovery of control and patient samples using the lots of IMMAGE BUF1. Some lots of BUF1 have been reported to cause high shifts in recovery of control or patient samples. Quality Control failures may result after replacing BUF1 with a different lot.

Action

Beckman Coulter will send a Product Corrective Action (PCA) letter on the week of April 19, 2010 to the affected customers. The letter provided the customers with an explanation of the problem and a work around.

Distribution

Worldwide: U.S., Canada, Australia, Austria, Belgium, China, Czech Republic, Denmark, France, Germany, Greece, Hong Kong, Iceland, Ireland, Italy, Japan, Liechtenstein, Macao, Malaysia, Netherlands, New Zealand, Norway, Oman, Poland, Qatar, Russian Federation, Singapore, Slovakia, South Africa, Spain, Sweden, Switzerland, Taiwan, Tunisia, Turkey and United Kingdom,