Medline Probe Cover Kits containing sterile Ultrasound Gel, Labeled as the following: a) CRANIOTOMY CDS-1, Model Number CDS982889I; b) JOINT INJECTION TRAY, Model Number DYNDH1722; c) LOWER EXTREMITY PACK-LF, Model Number DYNJ0604434X; d) ARTHROSCOPY PACK, Model Number DYNJ44322I; e) SHOULDER ARTHROSCOPY PACK, Model Number DYNJ44348D; f) SHOULDER ARTHROSCOPY PACK, Model Number DYNJ44348F; g) IR ARM PACK, Model Number DYNJ50712C; h) HYBRID PACK, Model Number DYNJ56907D; i) TUNNEL PACK, Model Number DYNJ58127B; j) CRANIOTOMY PACK, Model Number DYNJ61749A; k) CARPAL TUNNEL KIT, Model Number DYNJ63149A; l) ANGIOGRAPHY PACK, Model Number DYNJ65963A; m) ARTHROSCOPY PACK, Model Number DYNJ66402; n) ARTHROSCOPY PACK, Model Number DYNJ66402A; o) ARTHROSCOPY PACK, Model Number DYNJ66402C; p) SHOULDER ARTHROSCOPY PACK, Model Number DYNJ66404; q) SHOULDER ARTHROSCOPY PACK, Model Number DYNJ66404A; r) SHOULDER ARTHROSCOPY PACK, Model Number DYNJ66404B; s) ARTHROSCOPY PACK, Model Number DYNJ68997; t) MURRIETA PACK, Model Number DYNJ69312; u) MURRIETA PACK, Model Number DYNJ69312A; v) SHOULDER ARTHROSCOPY PACK, Model Number DYNJ82077; w) SCO ARTHROSCOPY, Model Number DYNJ906551A; x) US JOINT KIT, Model Number DYNJRA1944; y) HIP FRACTURE NERVE BLOCK, Model Number DYNJRA2085; z) ENS STERILE POUR IMPLANT-LF, Model Number OST011E; aa) ARTHROGRAM TRAY, Model Number PAIN1524C; bb) SHOULDER PACK-LF, Model Number PHS41748D; cc) LOWER EXTREMITY BLOCK PACK-RX, Model Number SPEC0204D; dd) UPPER EXTREMITY BLOCK PACK-RX, Model Number SPEC0205F;
Recall
- Recall Number
- Z-2191-2023
- Event Number
- 92466
- Firm
- MEDLINE INDUSTRIES, LP Northfield
- FEI Number
- 1417592
- Product Code
- OJH
- Status
- Open, Classified
- Root Cause
- Device Design
- Initiated
- May 18, 2023
- Posted
- July 20, 2023
- Address
- 3 Lakes Dr, Northfield, IL, 60093-2753
Description
Medline Probe Cover Kits containing sterile Ultrasound Gel, Labeled as the following: a) CRANIOTOMY CDS-1, Model Number CDS982889I; b) JOINT INJECTION TRAY, Model Number DYNDH1722; c) LOWER EXTREMITY PACK-LF, Model Number DYNJ0604434X; d) ARTHROSCOPY PACK, Model Number DYNJ44322I; e) SHOULDER ARTHROSCOPY PACK, Model Number DYNJ44348D; f) SHOULDER ARTHROSCOPY PACK, Model Number DYNJ44348F; g) IR ARM PACK, Model Number DYNJ50712C; h) HYBRID PACK, Model Number DYNJ56907D; i) TUNNEL PACK, Model Number DYNJ58127B; j) CRANIOTOMY PACK, Model Number DYNJ61749A; k) CARPAL TUNNEL KIT, Model Number DYNJ63149A; l) ANGIOGRAPHY PACK, Model Number DYNJ65963A; m) ARTHROSCOPY PACK, Model Number DYNJ66402; n) ARTHROSCOPY PACK, Model Number DYNJ66402A; o) ARTHROSCOPY PACK, Model Number DYNJ66402C; p) SHOULDER ARTHROSCOPY PACK, Model Number DYNJ66404; q) SHOULDER ARTHROSCOPY PACK, Model Number DYNJ66404A; r) SHOULDER ARTHROSCOPY PACK, Model Number DYNJ66404B; s) ARTHROSCOPY PACK, Model Number DYNJ68997; t) MURRIETA PACK, Model Number DYNJ69312; u) MURRIETA PACK, Model Number DYNJ69312A; v) SHOULDER ARTHROSCOPY PACK, Model Number DYNJ82077; w) SCO ARTHROSCOPY, Model Number DYNJ906551A; x) US JOINT KIT, Model Number DYNJRA1944; y) HIP FRACTURE NERVE BLOCK, Model Number DYNJRA2085; z) ENS STERILE POUR IMPLANT-LF, Model Number OST011E; aa) ARTHROGRAM TRAY, Model Number PAIN1524C; bb) SHOULDER PACK-LF, Model Number PHS41748D; cc) LOWER EXTREMITY BLOCK PACK-RX, Model Number SPEC0204D; dd) UPPER EXTREMITY BLOCK PACK-RX, Model Number SPEC0205F;
Probe covers for use during diagnostic ultrasound procedures may have an inadequate barrier at the seams.
Medline distributed their recall notices to their consignees beginning on 05/18/2023 by letter. The notices explained the problem, and the risk and requested the destruction of the products. Distributors were directed to notify their consignees, have the customers document and destroy any affected product, and include your customers quantities on the response form. Beginning on 05/23/2023, Medline issued another notice to a separate group of consignees requesting that the kits be over labeled. Distributors were directed to notify the customers and have them return any affected product to the distributor for collection and return to Medline. Questions: please contact the Recall Department at 866-359-1704
Worldwide Distribution: US (Nationwide) and OUS (foreign) United Arab Emirates, Singapore, Lebanon, China Qatar, Costa Rica, India, Canada, Panama & Saudi Arabia.
19940 units