FDA Recall Terminated

Part # 313.946, Low Profile Neuro Screwdriver Blade/MR Safe/MQC/56mm Medium, Lot # 5053374, mfg 7/27/2005 Product Usage: The DePuy Synthes MR Safe Instruments and MR conditional Instruments are used to assist with the insertion or removal of an implant.

Recall: Z-2190-2016 · Initiated May 31, 2016

Recall

Recall Number
Z-2190-2016
Event Number
74416
Firm
Synthes (USA) Products LLC
FEI Number
3008812560
Product Code
HWE
Status
Terminated
Root Cause
Labeling False and Misleading
Initiated
May 31, 2016
Posted
July 12, 2016
Terminated
February 2, 2017
Address
1301 Goshen Pkwy, West Chester, PA, 19380-5986

Description

Part # 313.946, Low Profile Neuro Screwdriver Blade/MR Safe/MQC/56mm Medium, Lot # 5053374, mfg 7/27/2005 Product Usage: The DePuy Synthes MR Safe Instruments and MR conditional Instruments are used to assist with the insertion or removal of an implant.

Reason

DePuy Synthes is voluntarily initiating this recall due to affected devices that were distributed, etched and or labeled with MR Safety information not meeting the current ASTM standard.

Action

DePuy Synthes mailed an URGENT MEDICAL DEVICE FIELD SAFETY NOTIFICATION letter to their customers (Hospital Personnel, including Imaging Department Personnel and DePuy Synthes Colleague) to inform them of the recall and actions to take. The letter identified the affected, problem and actions to be taken. The letter instructed customers not to use the affected Instruments which are labeled or etched MR Safe and MR Conditional during MRI scans. Customers were asked to review the information contained in the Field Safety Notification letter and complete the Verification Section located on page 3 of this notification. For questions call 610-719-5450 or contact your DePuy Synthes Sales Consultant.

Distribution

US Nationwide Distribution in the states of: AR,CA,CO,CT,DC,ID,IL,KY,LA,MA,MD,MN,MO,MS,MT,NY,PA and TX.

Quantity

26 units