FDA Recall Open, Classified

Philips Allura R8.2.x Systems. Software Verion: R8.2.x. 1. Allura Xper FD10, Model Number: 722026. 2. Allura Xper FD10/10, Model Number: 722027. 3. Allura Xper FD20, Model Number: 722028. 4. Allura Xper FD20/10, Model Number: 722029. 5. Allura Xper FD10 OR Table, Model Number: 722033. 6. Allura Xper FD20 OR Table, Model Number: 722035. 7. Allura Xper FD20/20, Model Number: 722038 8. Allura Xper FD20/20 OR Table, Model Number: 722039. 9. Allura Xper FD20/15, Model Number: 722058. 10. Allura Xper FD20/15 OR Table, Model Number: 722059.

Recall: Z-2189-2025 · Initiated June 30, 2025

Recall

Recall Number
Z-2189-2025
Event Number
97155
Firm
PHILIPS MEDICAL SYSTEMS NEDERLAND B.V. Veenpluis 4-6 Best Netherlands
FEI Number
3002807880
Product Code
OWB
Status
Open, Classified
Root Cause
Software design
Initiated
June 30, 2025
Posted
July 29, 2025

Description

Philips Allura R8.2.x Systems. Software Verion: R8.2.x. 1. Allura Xper FD10, Model Number: 722026. 2. Allura Xper FD10/10, Model Number: 722027. 3. Allura Xper FD20, Model Number: 722028. 4. Allura Xper FD20/10, Model Number: 722029. 5. Allura Xper FD10 OR Table, Model Number: 722033. 6. Allura Xper FD20 OR Table, Model Number: 722035. 7. Allura Xper FD20/20, Model Number: 722038 8. Allura Xper FD20/20 OR Table, Model Number: 722039. 9. Allura Xper FD20/15, Model Number: 722058. 10. Allura Xper FD20/15 OR Table, Model Number: 722059.

Reason

Potential for temporary loss of imaging (X-ray) functionality due to software issue.

Action

Philips notified consignees on 06/30/2025 via letter. Consignees were informed what the issue was and under what circumstances it can occur and how to identify affected units. Consignees were instructed to notify all relevant personnel, avoid rapidly pressing and releasing the footswitch pedals, if the X-ray functionality is not available during the issues described in the notification, follow specific outlined steps, and notify customers if the units have been transferred or further distributed. Additionally, consignees were requested to maintain the notification with the system until the software update is complete and to complete and return the response form.

Distribution

Worldwide - US Nationwide and the countries of Algeria, Argentina, Australia, Austria, Azerbaijan, Bangladesh, Belarus, Belgium, Brazil, Bulgaria, Canada, Cayman Islands, Chile, China, Colombia, Costa Rica, Croatia, Cuba, Cyprus, Czech Republic, Denmark, Ecuador, Egypt, Finland, France, French Polynesia, Georgia, Germany, Greece, Guinea, Honduras, Hong Kong, Hungary, India, Indonesia, Iran, Iraq, Ireland, Israel, Italy, Japan, Jordan, Kazakhstan, Kenya, Kosovo, Lao, Lebanon, Lithuania, Malaysia, Maldives, Malta, Martinique, Mexico, Mongolia, Morocco, Myanmar, Nepal, Netherlands, New Caledonia, Norway, Oman, Pakistan, Palestine, Panama, Papua New Guinea, Paraguay, Peru, Philippines, Poland, Portugal, Qatar, R¿union, Romania, Russian Fed., Saudi Arabia, Singapore, Slovakia, South Africa, South Korea, Spain, Sri Lanka, Sudan, Sweden, Switzerland, Syria, Taiwan, Tanzania, Thailand, Trinidad & Tobago, Tunisia, T¿rkiye, Ukraine, United Kingdom, Uruguay, United Arab Emirates, Uzbekistan, Vietnam, Yemen.

Quantity

4,508 units