FDA Recall Terminated

Hologic Explorer Bone Densitometers with Apex software versions 2.0, 2.1 and 2.2 is indicated for the estimate of bone mineral density, comparison of measured variables obtained from a given scan to a database of reference values, the estimation of fracture risk, vertebral deformity assessment, body composition analysis, and discrimination of bone from prosthetics.

Recall: Z-2187-2008 · Initiated June 27, 2008

Recall

Recall Number
Z-2187-2008
Event Number
48641
Firm
Hologic, Inc.
FEI Number
1221300
Product Code
KGI
Status
Terminated
Root Cause
Software design
Initiated
June 27, 2008
Posted
September 17, 2008
Terminated
October 25, 2010
Address
35 Crosby Dr, Bedford, MA, 01730-1450

Description

Hologic Explorer Bone Densitometers with Apex software versions 2.0, 2.1 and 2.2 is indicated for the estimate of bone mineral density, comparison of measured variables obtained from a given scan to a database of reference values, the estimation of fracture risk, vertebral deformity assessment, body composition analysis, and discrimination of bone from prosthetics.

Reason

Software densitometer readings for left hip and lumbar spine under certain conditions may be inaccurate.

Action

Hologic notified customers by registered letter titled Urgent: Medical Device Recall, dated 6/27/08, providing instructions to update the software to Version 2.3 and discard version APEX 2.0, 2.1 and 2.2 manuals. If you have questions or require assistance with the software installation, contact Hologic's Help Desk at 1-800-321-4659.

Distribution

Nationwide Foreign: Canada Algeria, AE, Australia, Belgium,China,France, Germany, Hungary, Hong Kong, Iceland, Iran, Ireland, Israel, Italy,Japan,Korea, Kuwait, Lithuania, Mexico, Myanmar, Netherlands, Pakistan, Peru, Russia, S. Africa, Saudi Arabia, Singapore, Slovak Rep., Switzerland, Syria, Taiwan, UK

Quantity

325 units