CellaVision DM Software versions 6.0.1 or 6.0.2 installed on the following products: CellaVision DM96, DM1200, DM9600 and DI-60 Product Usage: The devices are automated cell-locating devices. The devices automatically locates and presents images of blood cells on peripheral blood smears. The operator identifies and verifies the suggested classification of each cell according to type. The devices are intended to be used by skilled operators, trained in the use of the device and in recognition of blood cells

Recall: Z-2184-2018 · Initiated December 4, 2017

Recall

Recall Number
Z-2184-2018
Event Number
80137
Firm
Cellavision AB Forskningsbyn Ideon Scheelevagen 19a Lund Sweden
FEI Number
3003630693
Product Code
JOY
Status
Open, Classified
Root Cause
Software design
Initiated
December 4, 2017

Description

CellaVision DM Software versions 6.0.1 or 6.0.2 installed on the following products: CellaVision DM96, DM1200, DM9600 and DI-60 Product Usage: The devices are automated cell-locating devices. The devices automatically locates and presents images of blood cells on peripheral blood smears. The operator identifies and verifies the suggested classification of each cell according to type. The devices are intended to be used by skilled operators, trained in the use of the device and in recognition of blood cells

Reason

A software malfunction was found where WBC, RBC and PLT comments added after a slide is signed, are not sent to the LIS. This can only occur where customers process multiple slides per blood sample.

Action

CellaVision sent an Urgent Medical Device Recall letter dated December 4, 2017. The letter identified affected product, problem and the actions to be taken. Customers are requested to determine if they are affected or not. If affected they are requested to contact their local distributor for installing a patch. Customers are asked to confirm receipt of the notice by signing and dating the attached Acknowledgement Form and emailing a copy to their local distributor. For technical questions contact CelleVision ([email protected])

Distribution

US Nationwide in the states of FL, IL, NY, and NC

Quantity

Instruments: 224 Software: 99 (US)