FDA Recall Terminated

TEMPUS IC PROFESSIONAL Patient Monitor, Part No. 00-1002, Manufactured by: Remote Diagnostics Technologies Ltd., The Old Coach House, Farleigh Wallop, Hants, UK RG25 2HT. Indicated for 3 Lead ECG monitoring and 12 Lead ECG recording, non-invasive blood pressure (NIBP), respiration, end-tidal C02 (ETCO 2), pulse oximetry (SP02) and tympanic temperature.

Recall: Z-2183-2011 · Initiated March 25, 2011

Recall

Recall Number
Z-2183-2011
Event Number
58479
Firm
Remote Diagnostic Technologies Ltd. The Old Coach House, The Av. Farleigh Wallop, Basingstoke Hampshire United Kingdom
FEI Number
3003832357
Product Code
MWI
Status
Terminated
Root Cause
Component change control
Initiated
March 25, 2011
Posted
May 11, 2011
Terminated
September 16, 2011

Description

TEMPUS IC PROFESSIONAL Patient Monitor, Part No. 00-1002, Manufactured by: Remote Diagnostics Technologies Ltd., The Old Coach House, Farleigh Wallop, Hants, UK RG25 2HT. Indicated for 3 Lead ECG monitoring and 12 Lead ECG recording, non-invasive blood pressure (NIBP), respiration, end-tidal C02 (ETCO 2), pulse oximetry (SP02) and tympanic temperature.

Reason

It was discovered that when a 12 Lead ECG is recorded and monitoring is quickly restarted, the ECG can cease functioning.

Action

Remote Diagnostic Technologies, Ltd notified its consignees of the problem and the work-around via letter with the subject line "ECG Issue - Field Action Required" on 03/25/2011. The letter stated that customers can avoid the issues by viewing the recorded ECG for at least 13 seconds. Two diagrams were attached with the letter depicting the conditions under which the issue can occur. Customers can continue to use the device normally in all other respects. The letter states that the firm is working on a software modification which will permanently fix the issue and prevent it from occurring. Customers are to contact the Regulatory Affairs and Operations Manager if they have any questions.

Distribution

Nationwide Distribution -- KY, MD, VA, and FL.

Quantity

19 units