Dimension Vista Gentamicin Flex reagent cartridge, DV GENT, gentamicin enzyme immunoassay. Affected lots are 17135BC, 17150BF, 17159BD, 17178AC, 17269BD, and 17312BB.
Recall
- Recall Number
- Z-2182-2018
- Event Number
- 80009
- Firm
- Siemens Healthcare Diagnostics, Inc.
- FEI Number
- 2517506
- Product Code
- LCD
- Status
- Terminated
- Root Cause
- Under Investigation by firm
- Initiated
- April 19, 2018
- Terminated
- April 24, 2020
- Address
- 500 Gbc Dr Po Box 6101, Ms 514, Newark, DE, 19702-2466
Description
Dimension Vista Gentamicin Flex reagent cartridge, DV GENT, gentamicin enzyme immunoassay. Affected lots are 17135BC, 17150BF, 17159BD, 17178AC, 17269BD, and 17312BB.
The affected lots may exhibit inaccuracy for Quality Control and patient samples at the low end of the Analytical Measurement Range. The bias is due to the calibration curve not showing separation between the Level 1 and Level 2 calibrators. The negative bias observed for patient, QC, and calibrator samples, when using any of the listed lots, has the potential to impact the interpretation of gentamicin results during therapeutic monitoring at concentrations up to approximately 2.5 g/mL [5.4 mol/L].
On April 19, 2018, the firm sent an Urgent Medical Device Recall letter to customers. The letter identified the affected lots and advised customers to do the following: Inspect stock, discontinue use of and discard the Dimension and Dimension Vista GENT lots listed in Table 1. Please review this letter with your Medical Director. Review your inventory of these products to determine your laboratorys replacement needs and to provide information to Siemens for reporting to the regulatory authorities. Complete and return the Field Correction Effectiveness Check Form attached to this letter within 30 days. If you have received any complaints of illness or adverse events associated with the product listed in Table 1, immediately contact your local Siemens Customer Care Center or your local Siemens Technical Support Representative.
Nationwide distribution. Foreign distribution to Argentina, Australia, Brazil, Bahamas, Canada, Germany, India, Japan, Myanmar, New Zealand, Saudi Arabia, Taiwan, and South Africa.
20373