FDA Recall Terminated

GE Healthcare Revolution EVO X-ray system

Recall: Z-2181-2017 · Initiated May 19, 2017

Recall

Recall Number
Z-2181-2017
Event Number
77377
Firm
GE Healthcare, LLC
FEI Number
2126677
Product Code
JAK
Status
Terminated
Root Cause
Software design
Initiated
May 19, 2017
Terminated
July 29, 2018
Address
3000 N Grandview Blvd, Waukesha, WI, 53188-1615

Description

GE Healthcare Revolution EVO X-ray system

Reason

It was discovered that an issue has been identified that can result in a scan starting before the prescribed delay has completed.

Action

GE Healthcare proposed action to bring defect into compliance : 1. Inform users of the potential for this issue to occur and instruct users as to the actions to take in the event of an Emergency Stop actuation during prep-delay. 2. Provide an update to product labeling for users that includes this information. GE Healthcare will include the updated product labeling with all current production systems. 3. Make modifications during the next scheduled product update to prevent this "race" condition from occurring in newly produced systems. CDRH approves the CAP subject to the following conditions: 1. Notification of all dealers and purchasers is to be made within 15 working days of receipt of this letter in the manner specified in 21 CFR 1003.21 and 1003.22. This office and the Food and Drug Administration (FDA) district office coordinator noted below are to be included in the notification. 2. Any future customer notification letters to be sent must comply with the requirements of 21 CFR 1003.21. This includes, but is not limited to, the prescribed packaging and the statement that you will remedy the defect or bring the product into compliance at no charge. 3.You complete this recall by September 1, 2017. Report any failures regarding this recall immediately to the District Recall Coordinator. For further questions, please call (262) 513-4122.

Distribution

Worldwide Distribution - USA (nationwide) including Puerto Rico and to the countries of : Japan, Jordan, Kazakhstan, Korea ( Republic of) Kuwait, Latvia, Lebanon, Libya, Lithuania, Macedonia, Malaysia, Malta, Martinique, Mauritania, Mexico, Moldova, Morocco, Nambia, Netherlands, New Caledonia, New Zealand, Nigeria, Norway, Oman, Pakistan, Palestine, Panama, Paraguay, Peru, Philippines, Poland, Portugal, Qatar, Republic of Serbia, Reunion, Romania, Russia, Rwanda, Saudi Arabia, Singapore, Slovakia, Slovenia, South Africa, Spain Sweden, Switzerland, Syria, Taiwan, Thailand, Tunisia, Turkey, Ukraine, United Arab Emirates, United Kingdom, Uruguay, Uzbekistan, Venezuela, Vietam and Yemen.

Quantity

10,884