Digisonics, DigiView and ERS software release 3.6.4.4 through 3.6.5.2+, 2008 Digisonics, Inc., 3701 Kirby Drive, Houston TX 77098, 800-940-3240. DigiView is a digital image management and reporting system with an associated clinical database. It provides capture, review, and storage of images together with report creation and data basing.

Recall: Z-2180-2011 · Initiated January 20, 2009

Recall

Recall Number
Z-2180-2011
Event Number
58395
Firm
Digisonics, Inc
FEI Number
1626313
Product Code
LLZ
Status
Terminated
Root Cause
Device Design
Initiated
January 20, 2009
Posted
May 10, 2011
Terminated
May 31, 2011
Address
3701 Kirby Dr, Ste 930, Houston, TX, 77098-3922

Description

Digisonics, DigiView and ERS software release 3.6.4.4 through 3.6.5.2+, 2008 Digisonics, Inc., 3701 Kirby Drive, Houston TX 77098, 800-940-3240. DigiView is a digital image management and reporting system with an associated clinical database. It provides capture, review, and storage of images together with report creation and data basing.

Reason

Data from one patient's study had the potential for being put into another patient's report.

Action

Digisonics notified end users via e-mail January 20, 2009 and January 21, 2009, that Data from one patient's study was being put into another patient's report. The problem occurred when a System bug was found in DigiView Release 3.6.4.4, and higher versions in which under certain circumstances (specific workflow) a failure to switch to a patient's images occurs. Two software patch versions will be made available to sites on or after February 5, 2009, to correct the reported problem. If you have any further questions, please contact Digisonics Customer Support at (800 ) 940-3240.

Distribution

Worldwide Distribution -- USA including States of : OH, NC, LA, TX, CA, MO, NY, MD, MA, DC, IA, AZ, PA, MI, NC, FL, IN, SC, ID, MN, AR, OK, IL, CT, and countries of Canada, Lebanon, and New Zealand.

Quantity

70 units