St. Jude Medical, Engage TR Introducer, 6 F - ACT (2.25 mm), 7 cm length, .025" Max Guidewire O.D., Rx, Sterile EO, REF C408524, 100028093.
Recall
- Recall Number
- Z-2178-2010
- Event Number
- 56222
- Firm
- St. Jude Medical Cardiovascular Division
- FEI Number
- 2182269
- Product Code
- DYB
- Status
- Terminated
- Root Cause
- Process design
- Initiated
- June 28, 2010
- Posted
- August 12, 2010
- Terminated
- December 27, 2011
- Address
- 14901 Deveau Pl, Minnetonka, MN, 55345
Description
St. Jude Medical, Engage TR Introducer, 6 F - ACT (2.25 mm), 7 cm length, .025" Max Guidewire O.D., Rx, Sterile EO, REF C408524, 100028093.
Potential for a partial or complete separation of the Shaft (sheath) from the hub or a material break in the hub assembly just distal to the hemostasis seal. If either of these were to occur during use, it is likely that fluids would leak around the Introducer hub and strain relief.
Consignees were sent "Urgent Medical Device Recall Notice" dated June 24, 2010 from St. Jude Medical . The letter was addressed to Cath Lab Manger/ Risk Manager. The letter described the problem and the product involved. Advised consignees to discontinue use of the product and remove from inventory. Sales Representatives were also sent a St. Jude Medical Memorandum "Field Action Notice" dated June 24, 2010 and a reconciliation form. The memorandum described the problem and product. Advised Sales Rep to visit the affected accounts and to performed the assigned tasks. A St. Jude Medical sales representative will be contacting for removal and replacement of the affected product. For additional information, please contact St. Jude Medical at 651-491-4470.
Worldwide Distribution including USA, Italy, Denmark, France, Switzerland, England, Finland, Austria, Spain and Estonia.
5 OUS